8
COMPRESSOR USER MANUAL
866-DTE-INFO
dentalez.com
PN: 7717-004A
PLACEMENT
WARNING
DO NOT install on surfaces with more than a 5° incline.
CAUTION
The compressor is front heavy. When moving with
handles, be sure to hold it securely to prevent it from
tipping and possible dropping of the device, which could
result in damage to the compressor.
This device must not be stacked during transport and
storage and must be oriented right side up during
transport and storage.
The compressor unit should not be used adjacent to or stacked with other equipment. If adjacent or stacked use
is necessary, the compressor should be observed to verify normal operation in the configuration in which it will be
used.
Locate the compressor on a level hard surface capable of supporting the compressor weight. If a level surface
is not available, the compressor is equipped with height adjustable feet to level the device. Install four adjustable
rubber feet supplied to ensure unit stands firmly on the floor (see
SECTION III - INSTALLATION > FEET KIT
).
Floors should be wood, concrete or ceramic tile. It is highly recommended that a higher than 30% relative humidity
be established and the operator wear a ground strap before physical contact is attempted.
Compressor is for indoor use only, in a dust free, climate controlled room.
DO NOT install in a enclosed area where ambient temperature could exceed temperature specifications of below
35°F/02°C or above 100°F/38°C while the compressor is running.
Maintain minimum 12" clearance on all sides and on top of all compressors.
A source of conditioned fresh air is required for safe operation of this compressor. The air source must be located
outside of the utility room and away from potential air contaminants.
Do not locate this device or it’s air intake opening in an area that is prone to flooding or has previously flooded.
Water can damage components, reduce life expectancy of the compressor or cause serious personal injury.
TIPPING HAZARD
SECTION II - PREINSTALLATION