EPEX/Omniflex IV System Service Manual
4-12
Adjustments and Calibration
The following is an example of the Outputs screen.
The grey stripes mean that the function is not enabled/inactive. The red
stripes mean that the function is enabled/active.
The arrow in the lower middle area points to one of the five states
described below. Moving to the next state is accomplished by pressing the
State
button. The states are described in the following table.
State
Description
STANDBY
Sets state of the I/O when the Generator is in
standby or idle mode. Standby mode also defines the
state when the Generator is in fluoroscopic
hangover.
PREP
Sets state of the I/O when the Generator first enters
PREP mode.
GEN RDY
Sets state of the I/O when the Generator has
completed PREP mode and is ready to expose.
RAD EXP
Sets state of the I/O when the Generator starts a
radiographic exposure.
FLUORO EXP
Sets state of the I/O when the Generator starts a
fluoroscopic exposure.
Summary of Contents for EPEX Omniflex IV
Page 10: ......
Page 84: ...EPEX Omniflex IV System Service Manual 3 36 Installing the System Hardware...
Page 138: ...EPEX Omniflex IV System Service Manual 4 54 Adjustments and Calibration...
Page 146: ......
Page 182: ......
Page 184: ......
Page 186: ......
Page 188: ......
Page 190: ......
Page 192: ......
Page 194: ......
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 210: ......
Page 212: ......
Page 223: ...EPEX Omniflex System Service Manual Field Replaceable Units D 11...
Page 224: ......
Page 226: ......