EPEX/Omniflex IV System Service Manual
Adjustments and Calibration
4-39
If you want to discard the accumulated calibrations and start over,
click on the
Restart Calibration Sequence
button.
Step
Action
5
Change the technique settings, as defined in your Calibration Test
panel.
6
Remove the grid.
7
Take an exposure.
The progress of the calibration displays in the calibration panel.
If the calibration is successful,
a
1
displays in the calibration panel.
If the calibration is not successful,
adjust the technique settings and
take another exposure.
8
Click the
Accumulate Calibration
button.
9
Repeat steps 7 and 8 until you have accumulated four calibrations.
10
When you are finished the calibration sequence, click the
End
Calibration
button.
Summary of Contents for EPEX Omniflex IV
Page 10: ......
Page 84: ...EPEX Omniflex IV System Service Manual 3 36 Installing the System Hardware...
Page 138: ...EPEX Omniflex IV System Service Manual 4 54 Adjustments and Calibration...
Page 146: ......
Page 182: ......
Page 184: ......
Page 186: ......
Page 188: ......
Page 190: ......
Page 192: ......
Page 194: ......
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 210: ......
Page 212: ......
Page 223: ...EPEX Omniflex System Service Manual Field Replaceable Units D 11...
Page 224: ......
Page 226: ......