![Del Medical EPEX Omniflex IV Service Manual Download Page 143](http://html1.mh-extra.com/html/del-medical/epex-omniflex-iv/epex-omniflex-iv_service-manual_2465595143.webp)
EPEX/Omniflex IV System Service Manual
Preventive Maintenance
5-5
Annual Preventive Maintenance
EPEX Omniflex System Annual.cdr
Annual Preventive Maintenance Checklist
System:
Comments
Inspection
Open the Generator cabinet and examine the unit
for any visible damage: missing or loose ground
connections, oil leaks, damaged cables, and so on.
Ensure that there are no obstructions blocking any
of the ventilation holes or louvers on the Generator
cabinet.
Inspect ground, handswitch, operator contols,
control cables, and strain reliefs for wear and
pinching.
Check U-arm cable drape; inspect for cable sheath
wear, pinching, or excessive bending.
Check sheathing on cables that exit the U-arm and
enter the ceiling for pinching, cuts, fraying, or
excessive bending.
Check tightness of all electrical cable connections.
Check for proper operation of cooling fans; check
that blades are clean.
Check markings on controls and positioner for
legibility.
Summary of Contents for EPEX Omniflex IV
Page 10: ......
Page 84: ...EPEX Omniflex IV System Service Manual 3 36 Installing the System Hardware...
Page 138: ...EPEX Omniflex IV System Service Manual 4 54 Adjustments and Calibration...
Page 146: ......
Page 182: ......
Page 184: ......
Page 186: ......
Page 188: ......
Page 190: ......
Page 192: ......
Page 194: ......
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 210: ......
Page 212: ......
Page 223: ...EPEX Omniflex System Service Manual Field Replaceable Units D 11...
Page 224: ......
Page 226: ......