EPEX/Omniflex IV System Service Manual
Installing the System Hardware
3-23
Figure 3-17.
Inserting key switch
wires into collimator
power supply box
Figure 3-18.
Attaching the wires
5
Using a Phillips screwdriver and the screws (P/N 200-0029)
provided, attach the bottom cover to the generator.
6
Thread the wires from the key switch through the hole in the bottom
of the collimator power supply box, as in Figure 3-17.
7
Using a small slotted screwdriver, attach the two red wires from the
keyswitch to the terminals to which the red wires are attached
(numbers 2 and 3) along the bottom row of terminals. Remove the
red wires that were previously attached there and tighten the
terminal screws (see Figure 3-18).