EPEX/Omniflex IV System Service Manual
4-38
Adjustments and Calibration
Calibrating the DirectRay Detector
The DirectRay Detector should be calibrated weekly at the routine chest
orientation of 72 in. HSID. In addition, if the daily flat field exposure
procedure produces image tones that are not uniform or are otherwise
unsatisfactory, the DirectRay Detector should be calibrated.
To calibrate the DirectRay Detector
:
Step
Action
1
Ensure that the DirectRay Detector has been powered on for at least
one hour so that it has reached thermal equilibrium.
2
Ensure everyone is out of the X-ray room.
3
In the DirectRay Console application’s main window, close any
open studies and ensure that no patient is selected.
4
From the
Admin
menu, choose
Calibrate
.
The Calibration Test panel displays. This panel includes instructions
for how to set the technique settings and take exposures for the
calibration.
Note: Refer to the DirectRay Operator Console Service Manual for
specific images of the Calibration Test panel and other screens.
Summary of Contents for EPEX Omniflex IV
Page 10: ......
Page 84: ...EPEX Omniflex IV System Service Manual 3 36 Installing the System Hardware...
Page 138: ...EPEX Omniflex IV System Service Manual 4 54 Adjustments and Calibration...
Page 146: ......
Page 182: ......
Page 184: ......
Page 186: ......
Page 188: ......
Page 190: ......
Page 192: ......
Page 194: ......
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 210: ......
Page 212: ......
Page 223: ...EPEX Omniflex System Service Manual Field Replaceable Units D 11...
Page 224: ......
Page 226: ......