EPEX/Omniflex IV System Service Manual
Diagnostics and Troubleshooting
7-13
B0
Sync Pulse Error
Press
RESET
(located in the radiography
control group) and try the exposure again.
B1
No Digital Input
• Exposure request timed out
• Bucky contacts no closing
• No exposure taken
1. Exposure never started.
2. Exposure not taken during 8-second
time available.
3. Time out flag.
B2
Abort Exposure
No mAs
Handswitch let off too soon.
B3
X-ray Switch Jammed
Check the handswitch or wiring.
B5
mAs Not Reached and Time
was Stopped
• Too low mA
• Filament out of calibration
• Time exceeded 130% of the
expected mAs
1. Press
RESET
(located in the
radiography control group) and try the
exposure again.
2. Suspect miscalibration.
2B
Portrait/Landscape
Position Fault
Verify Collimator is On.
Table 7-2.
System-Level Error Codes
Error
Error Type
Possible Cause
Recommended Action
Summary of Contents for EPEX Omniflex IV
Page 10: ......
Page 84: ...EPEX Omniflex IV System Service Manual 3 36 Installing the System Hardware...
Page 138: ...EPEX Omniflex IV System Service Manual 4 54 Adjustments and Calibration...
Page 146: ......
Page 182: ......
Page 184: ......
Page 186: ......
Page 188: ......
Page 190: ......
Page 192: ......
Page 194: ......
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 210: ......
Page 212: ......
Page 223: ...EPEX Omniflex System Service Manual Field Replaceable Units D 11...
Page 224: ......
Page 226: ......