EPEX/Omniflex IV System Service Manual
4-30
Adjustments and Calibration
Aligning the Collimator Light Field to the X-ray Field
Note: If the light field to the
X-ray is too small, refer to the
Collimator service manual
that came with the system.
To align the Collimator light field to the X-ray field:
Figure 4-6.
Example of Alignment
at 1%
Step
Action
1
Set SID to 100 cm.
2
Adjust the Collimator shutters so that the edges of the light field
coincide with the rectangular outline of the Collimator test tool.
Note: Ignore the Collimator crosshair.
3
Initiate an exposure at approximately 60 kVp and 8 mAs.
4
Accept the image.
5
View the image in DRIVEL.
If the X-ray field falls just within the image of the rectangular
frame,
there is good alignment. Refer to Figure 4-6.
If an edge of the X-ray field falls on the first spot,
±
1 cm on either
side of the line,
the edges of the X-ray and light fields are
misaligned by 1% of the distance between the X-ray source and the
Bucky.
An edge falling on the second spot,
±
2 cm, indicates an error of 2%
at 100 cm. The maximum allowed misalignment is 1% of the source
to image distance.
Summary of Contents for EPEX Omniflex IV
Page 10: ......
Page 84: ...EPEX Omniflex IV System Service Manual 3 36 Installing the System Hardware...
Page 138: ...EPEX Omniflex IV System Service Manual 4 54 Adjustments and Calibration...
Page 146: ......
Page 182: ......
Page 184: ......
Page 186: ......
Page 188: ......
Page 190: ......
Page 192: ......
Page 194: ......
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 210: ......
Page 212: ......
Page 223: ...EPEX Omniflex System Service Manual Field Replaceable Units D 11...
Page 224: ......
Page 226: ......