EPEX/Omniflex IV System Service Manual
Diagnostics and Troubleshooting
7-11
A3
Bucky Fault
Drive motor not moving
1. Verify U-arm J1 connector has power.
2. Deselect Bucky and press
RESET
(located in the radiography control
group).
3. Check the Bucky Enable and Bucky
Return signals.
4. Verify J1-12 and J1-13 close in Bucky.
5. Verify Generator Panel N1 J4-12 and 13
have signals.
6. Verify System Controller Motherboard
J20-1 has signal and goes to 8 V gnd.
Verify J20-2 and J39-22 have signals.
7. Verify Generator Panel M1 J1-22 has
signal.
8. Verify WAMI J1-22 has signal.
A4
(Reserved)
A5
Communication Fault
• Communication fault
between the DirectRay
Console and the Generator
• Tube arcing
• Handswitch prematurely
released
1. Released handswitch too early.
2. Check the prep expose_ready signals
going to the Generator, and the
prep_ready and xray_ready signals from
the Generator.
A6
kV Error
• mA problems
• Filament temperature too
high
1. Suspect miscalibration.
2. Check the kV feedback on the HF Power
Board.
3. Scope hV transformer; check waveform.
4. Verify TP2 and TP3 for mA and kV.
A7
Filament Error
• Overcurrent
• Tube arcing
• Open circuit; no Filament
• Cable problem
• Cathode fault from IGBT
drivers
Review Filament circuit.
A8
mA Overload Error
• Bad plug
• mA out of bounds for
exposure
• Filament circuit fault
1. Verify mA requested is within bounds.
2. Review Filament circuit.
3. Check plugs.
Table 7-2.
System-Level Error Codes
Error
Error Type
Possible Cause
Recommended Action
Summary of Contents for EPEX Omniflex IV
Page 10: ......
Page 84: ...EPEX Omniflex IV System Service Manual 3 36 Installing the System Hardware...
Page 138: ...EPEX Omniflex IV System Service Manual 4 54 Adjustments and Calibration...
Page 146: ......
Page 182: ......
Page 184: ......
Page 186: ......
Page 188: ......
Page 190: ......
Page 192: ......
Page 194: ......
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 210: ......
Page 212: ......
Page 223: ...EPEX Omniflex System Service Manual Field Replaceable Units D 11...
Page 224: ......
Page 226: ......