vPad-Mini OPERATORS MANUAL
Figure 2
If there is a problem during initialization of the hardware, an Alert Dialog is shown (
Figure 2
).
Fault messages which may be shown by the dialog are:
Error Message
Shown if...
Analyzer electronics off-line !
Check AC supply input.
Attempt to connect with
analyzer via DACOM times out
Analyzer will not initialize !
Service is required.
No response to hardware
initialization command, or
No response to hardware idle
command
Unable to calibrate voltmeter !
Service is required.
No response to voltmeter
calibration command
Voltmeter offset is too high !
Service is required.
Response to command exceeds
maximum # of counts
The
Close
button dismisses the dialog and retries making a communication connection. If the hardware
fault cannot be corrected, operation loops indefinitely on the dialog until the phone is switched off.
Installation/Chapter 2
#
Page 4