vPad-Mini OPERATORS MANUAL
Figure 47
Page 3,
Figure 47
, allows changes to the Protective Earth limits. The normal, default limit for the Protective
Earth measurement is indicated in yellow. Alternative limits that may be allowed in the standard under
certain circumstances, or that may have been commonly used previously are also provided. The Mains
criteria may also be changed, if desired, for this device test.
If Semi-Auto(mated) mode is enabled, the vPad-Mini will stop and wait for instructions at any point where
changing the Mains test condition is about to happen, or where user input is required.
If no more changes to the protocol are required, touch the Run icon,
, to start the test. Otherwise,
touch the Next page icon,
to continue changes, or the back icon
to go back one page.
Page 4,
Figure 48
, provides the final test configuration options: Protective Earth, Mains Voltage and Device
Current. When selections are complete, you can touch the Run icon,
, to start the test. To save this
(new) test protocol as a (named) Sequence for future use, press the Save icon,
. Note that this named
sequence will be recorded for the current test report. There is also an opportunity when saving the test
result to save it as a named Sequence.
Operation/Chapter 4
#
Page 33
Summary of Contents for vPad-Mini
Page 1: ...Operating Manual MN 125b vPad Mini vPad Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...vPad Mini OPERATORS MANUAL Table of Contents Page iv ...
Page 10: ...vPad Mini OPERATORS MANUAL Overview Chapter 1 Page 2 ...
Page 52: ...vPad Mini OPERATORS MANUAL Operation Chapter 4 Page 44 ...
Page 54: ...vPad Mini OPERATORS MANUAL ...