Datex-Ohmeda S/5 Anesthesia Monitor
Document No. 800 1005-1
2
1
SPECIFICATIONS
1.1
General specifications
Module size, W
×
D
×
H
135
×
410
×
135 mm/5.3
×
15.0
×
5.3 in
Module weight
6 kg/13 lbs.
1.2
Typical performance
Sampling rate
200 ml/min nominal (180...220 ml/min)
Display update rate
breath-by-breath
Automatic compensation for pressure, CO
2
-N
2
O, and CO
2
-O
2
collision broadening effect.
Warm-up time 3 min for operation, 30 min for full specifications.
Auto-zeroing is performed at start-up, after 5 min + 5 min + 5 min + 15 min + 15 min + 15 min,
and after that every 60 min at regular intervals.
1.2.1
CO
2
Measurement range
0 to 10 %, (0 to 10 kPa), (0 to 76 mmHg)
Extended range
10 to 15 %, (10 to 15 kPa), (76 to 114 mmHg) (unspecified)
If CO
2
concentration is below 0.1 %, 0.0 % is displayed.
1.2.2
Respiration rate
Breath detection
1 % change in CO
2
level
Measurement range
4 to 60 breaths/min
1.2.3
O
2
Measurement range
0 to 100 % O
2
1.2.4
N
2
O
Measurement range
0 to 100 % N
2
O
1.2.5
Hal, Iso, Enf
Measurement range
0 to 5 %
Extended range
5 to 15 % (unspecified)
1.2.6
Sev
Measurement range
0 to 8 %
Extended range
8 to 15 % (unspecified)
Summary of Contents for S/5 B-GAS
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 iv ...
Page 25: ...S 5 Airway modules Document No 800 1005 1 19 Figure 11 ACX measuring board block diagram ...
Page 77: ...APPENDIX A Service check form S 5 Airway modules Document No 800 1005 1 71 APPENDIX A ...
Page 78: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 72 ...
Page 82: ...Datex Ohmeda S 5 Anesthesia Monitor A 4 3 Document No 800 1005 1 ...
Page 83: ......