Datex-Ohmeda S/5 Anesthesia Monitor
Document No. 800 1005-1
54
4.8
Error messages
Message
Explanation
Occlusion
The sample tube inside or outside the monitor is blocked or water trap is
occluded. If occlusion persists, measured gas values disappear.
Air leak
- the water trap is not connected
- the gas outlet is blocked
- there is a leak in the sampling line inside the module. If air leak persists
measured gas values disappear.
Replace trap
Indicates residue build-up on the water trap membrane. This decreases air
flow.
Zero valve error
Opening the valve does not change working pressure enough.
Gas calibration is not available during the first 5
minutes/during occlusion/during air leak
Calibration not allowed during the first 5 minutes after power up and in
mentioned situations.
Select agent
No agent selected.
Continuous occlusion. Check sampling line and water trap.
Occlusion over 40 seconds.
Air leak detected. Check water trap and sample gas out-flow.
Press normal screen to continue.
Air leak over 40 seconds.
CO2
Zero error
Unsuccessful zeroing
Unstable
Unsuccessful calibration
CO2 over scale
CO2 signal exceeds the maximum waveform area
O2
O2 zero error
Unsuccessful zeroing
O2 over scale
O2 signal exceeds the maximum waveform area
O2 Unstable
Unsuccessful calibration
N2O
N2O zero error
Unsuccessful zeroing
N2O Unstable
Unsuccessful calibration
Ane agents
AA zero error
Zero error
Unsuccessful zeroing
AA unstable
Unstable
Unsuccessful calibration
AA over scale
AA signal exceeds the maximum waveform area
Menu messages during calibration
Zero error
Unsuccessful zeroing
Adjust
Calibration gas accepted and monitor is ready for adjusting the gas values
to match the calibration gas concentration
Unstable
Unsuccessful calibration
Summary of Contents for S/5 B-GAS
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 iv ...
Page 25: ...S 5 Airway modules Document No 800 1005 1 19 Figure 11 ACX measuring board block diagram ...
Page 77: ...APPENDIX A Service check form S 5 Airway modules Document No 800 1005 1 71 APPENDIX A ...
Page 78: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 72 ...
Page 82: ...Datex Ohmeda S 5 Anesthesia Monitor A 4 3 Document No 800 1005 1 ...
Page 83: ......