S/5 Airway modules
Document No. 800 1005-1
63
5.3.2
Temp & Hum service menu
If circumstances noticeably differ from normal, or additional accuracy is required, the use of Temp
& Humidity menu may be advisable.
Especially small errors in tidal values may indicate that temperature and humidity settings of the
monitor differ too much from the used system.
Room Temperature and Humidity
These are needed only in calibration procedure.
Insp Temperature
The setting regarding the temperature of inspired gas. The value is used in calculations. Change if
necessary.
Insp Humidity
The setting regarding the humidity of inspired gas. The value is used in calculations. Change if
necessary.
Exp Temperature
The setting regarding the temperature of expired gas. The value is used in calculations. Change if
necessary.
Exp Humidity
The setting regarding the humidity of expired gas. The value is used in calculations. Change if
necessary.