S/5 Airway modules
Document No. 800 1005-1
35
8.
If the module contains a membrane key on the front panel, press the key at least for one
second and check that it is identified, i.e. the text for Button changes from OFF to ON in the
menu.
?
•
Select Halothane as anesthetic agent by first selecting AGENTS from the ACX service menu:
Agents - Select Agent - Hal - Previous Menu
9.
Check that the Calib zero -value for N
2
O is less than 61000.
If the value exceeds the limit, it indicates bad contamination in the ACX measuring
chamber. The measuring chamber can be cleaned according to the special cleaning
instructions found in the chapter Cleaning the measuring chamber of ACX measuring unit.
However, if cleaning does not help, the whole ACX measuring unit should be exchanged.
NOTE: If the ACX measuring chamber was cleaned or the unit was replaced, then also the
tubing between the D-fend and the ACX measuring unit, including the zero valve, should be
replaced. The mentioned parts should not be cleaned.
NOTE: With monitor software S-___95 and S-___96 the
Calib zero
-value for N
2
O is always
shown as a negative value. The correct value can be calculated by adding the shown value
to the value 65536.
?
10.
Check that the
Ambient
-value corresponds with the current ambient pressure (±20 mmHg).
?
11.
Perform the steam test for the special tubes (Nafion), or replace the tubes.
NOTE: The special tubes are recommended to be replaced annually.
?
12.
Check the CO
2
absorber.
Keep the tip of the sampling line away from you and let the monitor draw in room air. Check
the “Insp CO
2
” -value from the ACX service menu. If the value is less than 4, replace the CO
2
absorber.
?
13.
Check the zero valve.
Feed calibration gas into the sampling line and check that the gas readings in the service
menu correspond with the gas values on the calibration gas bottle. Keep feeding gas and
activate the zero valve from the menu. The O
2
reading should drop back near 21 %, the
other gas readings near 0 %. If the readings did not drop, replace the zero valve.
Summary of Contents for S/5 B-GAS
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 iv ...
Page 25: ...S 5 Airway modules Document No 800 1005 1 19 Figure 11 ACX measuring board block diagram ...
Page 77: ...APPENDIX A Service check form S 5 Airway modules Document No 800 1005 1 71 APPENDIX A ...
Page 78: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 72 ...
Page 82: ...Datex Ohmeda S 5 Anesthesia Monitor A 4 3 Document No 800 1005 1 ...
Page 83: ......