![Datex-Ohmeda S/5 B-GAS Technical Reference Manual Download Page 68](http://html.mh-extra.com/html/datex-ohmeda/s-5-b-gas/s-5-b-gas_technical-reference-manual_513814068.webp)
Datex-Ohmeda S/5 Anesthesia Monitor
Document No. 800 1005-1
62
5.3.1
Flow calibration
Flow calibration
PVX board is calibrated at the factory and due to the board’s design calibration is not regularly
needed. The calibration data is saved into the board’s EEPROM memory and if the software EPROM
of the board is changed the calibration must be performed. It is recommended to perform the
calibration both with adult values using the D-lite, and with pediatric values using Pedi-lite.
1.
Connect a spirometry tube with the D-lite sensor to the airway module. To improve the
accuracy, the endotracheal tube and all accessories which are in normally use should be
attached also during the calibration.
2.
Enter service menu (
Monitor Setup
- Install Service - Service - Parameters
).
3.
Enter the PVX menu. After the flow is zeroed (‘Zero OK’ message displayed) attach
calibration pump or spirometry tester to the flow sensor (D-lite or Pedi-lite). Select the
sensor type.
4.
Set the calibration volume for adult to 1000 ml and for pediatric to 300 ml.
5.
Work on the calibration pump slowly, approximately 1 pump in 5 seconds, pump until
‘adjust’-message appears. If you use the spirometry tester, perform the calibration
according to the tester instructions.
6.
Adjust the reading to match the calibration volume (1000 ml for the D-lite and 300 ml for
the Pedi-lite).
NOTE: The last calibration dates are saved into the main CPU board memories.
Summary of Contents for S/5 B-GAS
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 iv ...
Page 25: ...S 5 Airway modules Document No 800 1005 1 19 Figure 11 ACX measuring board block diagram ...
Page 77: ...APPENDIX A Service check form S 5 Airway modules Document No 800 1005 1 71 APPENDIX A ...
Page 78: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 72 ...
Page 82: ...Datex Ohmeda S 5 Anesthesia Monitor A 4 3 Document No 800 1005 1 ...
Page 83: ......