Perform gas calibration:
AIRWAY GAS - GAS CALIBRATION
NOTE: For maximum accuracy, a warm-up time of 30 minutes is recommended.
NOTE: If the module contains the agent identification unit, the ASX-100 or ASX-200, keep
feeding gas at least 15 seconds after the message ‘Adjust’ appears in the menu. This way
the agent identification unit has enough time for calibration.
?
•
Enter the ACX service menu:
Monitor Setup
- Install/Service
(password 16-4-34)
- Service
(password 26-23-8) -
Parameters - Gas Unit - ACX
•
Select Halothane as anesthetic agent:
Agents - Select Agent - Hal - Previous Menu
20.
Perform the fall time measurement in the ACX service menu.
Check that the measured fall times are within the ranges that are given in the Technical
Reference Manual.
NOTE: The fall time measurement can be performed only with module software 884295 or
885388.
?
21.
Perform the noise measurement.
Check that the measured noise values are within the ranges that are given in the Technical
Reference Manual.
NOTE: The noise measurement can be performed only with module software 884295 or
885388.
?