S/5 Airway modules
Document No. 800 1005-1
33
All modules
•
Remove the airway module case, the top protection cover and the bronze plate from the
side of the O
2
sensor, if installed.
•
Detach the ACX measuring board and the PVX board with the support plate, if installed.
NOTE: Wear a static control wrist strap when handling PC boards. Electrostatic discharge may
damage components on the board.
1.
Check internal parts:
- screws are tightened properly
- cables are connected properly
- all IC’s that are on sockets are attached properly
- tubes are not pinched and there are no sharp bends on them
- tubes are connected properly
- there are no loose objects inside the module
NOTE: Make sure that none of the tubes is in contact with the sampling pump or the O
2
sensor.
?
2.
Check external parts:
- the module case is intact
- the four rubber pads under the frame are all in place
- the metal rear panel is intact
- the equipotential tap and the sample gas out connector are tightened properly
- the block screws for the gas interface cable are in place and are tightened properly
- the D-fend latch is functioning properly
?
•
Install the ACX measuring board.
•
Detach the D-fend.
3.
Check the condition of the rubber O-rings on the metal D-fend connectors, located inside
the module front cover.
If necessary, detach the connectors by first disconnecting the tubes, then removing the
locking rings from the back of the front cover.
NOTE: The O-rings are recommended to be replaced annually.
?
4.
Check the OM ref. filter (order code 86901) visually, if installed.
NOTE: The OM ref. filter is recommended to be replaced annually.
If the module does not contain the OM ref. filter, install it to prevent dirt entering the O
2
sensor reference channel:
Summary of Contents for S/5 B-GAS
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 iv ...
Page 25: ...S 5 Airway modules Document No 800 1005 1 19 Figure 11 ACX measuring board block diagram ...
Page 77: ...APPENDIX A Service check form S 5 Airway modules Document No 800 1005 1 71 APPENDIX A ...
Page 78: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 72 ...
Page 82: ...Datex Ohmeda S 5 Anesthesia Monitor A 4 3 Document No 800 1005 1 ...
Page 83: ......