Switch the monitor off, disconnect the gas interface cable and reassemble the module.
NOTE: When reassembling the module, make sure that the tubes and cables are not
pinched between the PC boards and covers.
29.
Clean, or replace the airway module fan filter.
?
30.
Perform electrical safety check and leakage current test.
?
•
Reconnect the gas interface cable and sampling line, switch the monitor on and wait until
the message ‘Calibrating gas sensor’ disappears from the screen.
31.
Block the tip of the sampling line with your finger and check that the message ‘Sample line
blocked’ appears onto the monitor screen within 30 seconds.
?
32.
Detach the D-fend and check that the message ‘Check D-fend’ appears onto the screen
within 30 seconds.
Reattach the D-fend.
?
•
Simulate at least 5 breaths by feeding calibration gas into the sampling line.
Check that the shown gas information is correct.
33.
Check that the monitor activates the APNEA -alarm within 30 seconds after you have
stopped feeding gas.
?
34.
Switch the monitor off, disconnect the gas interface cable and clean the module.
?
•
Fill in all necessary documents.
Summary of Contents for S/5 B-GAS
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 iv ...
Page 25: ...S 5 Airway modules Document No 800 1005 1 19 Figure 11 ACX measuring board block diagram ...
Page 77: ...APPENDIX A Service check form S 5 Airway modules Document No 800 1005 1 71 APPENDIX A ...
Page 78: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 72 ...
Page 82: ...Datex Ohmeda S 5 Anesthesia Monitor A 4 3 Document No 800 1005 1 ...
Page 83: ......