S/5 Airway modules
Document No. 800 1005-1
57
Serial Comm(unication)
Serial Communication indicates a state of serial communication between the module processor
and a measuring unit.
GAS:
FFFF Continuously
ACX:
FFFF Continuously
ASX:
Value is for factory use only.
PVX:
Value is for factory use only.
Rep status
Rep status is a four-digit number, where all digits, abcd , can have different values.
Gas rep status:
a
0
No sevoflurane or desflurane measurement available
3
ACX can measure sevoflurane and desflurane
b
0
No gas measurements available
F
CO2, O2 N2O and AA measurements available
3
CO2 and O2 available
c
0
No ACX, ASX, nor PVX board running
1
ACX board running
3
ACX and ASX board running
5
ACX and PVX board running
7
ACX, ASX and PVX board running
d
0
Normal operation state
1
Occlusion
2
Air leak
4
Other sampling system error
8
Replace trap
ACX rep status:
a
empty
Normal operation state
1
ACX initialization
b
empty
Normal operation state
2
Occlusion
4
Air leak
c
C
Normal operation state
others
values used in manufacturer's testing
d
0
Normal operation state
others
values used in manufacture's testing
ACX rep status
FFFF continuously
Summary of Contents for S/5 B-GAS
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 iv ...
Page 25: ...S 5 Airway modules Document No 800 1005 1 19 Figure 11 ACX measuring board block diagram ...
Page 77: ...APPENDIX A Service check form S 5 Airway modules Document No 800 1005 1 71 APPENDIX A ...
Page 78: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 72 ...
Page 82: ...Datex Ohmeda S 5 Anesthesia Monitor A 4 3 Document No 800 1005 1 ...
Page 83: ......