Before every patient
Inspect the system
Before each case, perform the following anesthesia system checks.
w
W
W
W
WA
A
A
AR
R
R
RN
N
N
NIIIIN
N
N
NG
G
G
G
Make sure that the breathing circuit is correctly connected and not
damaged.
w
C
C
C
CA
A
A
AU
U
U
UTTTTIIIIO
O
O
ON
N
N
N
The total weight on each accessory shelf must be less than 23 kg.
Systems without accessory shelves have a weight limit on the top
surface of 23 kg.
Make sure that:
1.
The equipment is not damaged.
2.
All components are correctly attached.
3.
The breathing circuit is correctly connected, not damaged, and contains
sufficient absorbent.
4.
The vaporizers are locked in position and contain sufficient agent.
5.
Pipeline gas supplies are connected and the pressures are correct.
6.
Cylinder valves are closed on models with cylinder supplies.
w
w
w
w
W
W
W
WA
A
A
AR
R
R
RN
N
N
NIIIIN
N
N
NG
G
G
G
Do not leave gas cylinder valves open if the pipeline supply is in
use. Cylinder supplies could be depleted, leaving an insufficient
reserve supply in case of pipeline failure.
7.
Models with cylinder supplies have a cylinder wrench attached to the
system.
8.
The necessary emergency equipment is available and in good condition.
9.
Equipment for airway maintenance, tracheal intubation, and IV
administration is available and in good condition.
10. Applicable anesthetic and emergency drugs are available.
Summary of Contents for 2 Vap. Trolley
Page 1: ...Aestiva 5 Operation Manual Part 1 Software Revision 4 X System Controls Operation Checkout ...
Page 6: ...Aestiva iv 1006 0938 000 iv ...
Page 14: ...Aestiva 1 8 1006 0938 000 1 8 ...
Page 36: ...Aestiva 2 22 1006 0938 000 ...
Page 76: ...Aestiva 4 4 1006 0938 000 ...
Page 102: ...Aestiva I 4 1006 0938 000 ...