All but two of the controls for the ventilator are located on the Ventilation/
Monitoring display. The two controls are:
• The system switch, which powers the ventilator.
• The Bag/Vent switch, which starts and stops mechanical ventilation.
Item
Description
Menu key
Shows the main menu.
Alarm silence key
and indicator
Silences most alarms for 120 seconds.
Pushing the key when no alarm is active pre-silences
low and medium priority alarms, except Minimum
Monitoring, for 90 seconds.
The "No O2 Pressure" alarm cannot be silenced.
Volume alarm key
and status
Turns volume alarms on and off.
AB.90.036
Remaining
silence time
120
AB.29.004
Vol Alarms Off
Vol Alarms On
Vol Alarms Off