Aestiva
2-6
1006-0938-000
.
Item
Description
Bag/Vent switch
When you first turn on the system, mechanical ventilation is always off. To start mechanical
ventilation, move the switch from Bag to Vent.
APL valve
Limits breathing system pressure during manual ventilation. The scale shows approximate
pressures. Above 30 cmH
2
O, you will feel clicks as the knob turns.
Bag arm
One arm is adjustable (push in and turn). The other is not.
Mechanical ventilation On (gas to bellows)
Mechanical ventilation Off (gas to bag arm)
AB.23.059
AB.23.060
~0 cmH
2
O
~20 cmH
2
O
20
30
70
5
20
30
70
5
Increase
AB.23.064
Adjustable
Not adjustable
AB.23.023
AB.23.024
,