Appendix - Preoperative Tests
1006-0938-000
A-9
•If the O
2
flow also decreases more than 1 L/min with a different
vaporizer, the malfunction is in the Aestiva. Do not use the Aestiva
system until it is serviced.
5.
Complete steps 3 and 4 for each vaporizer.
Power failure test
1.
Unplug the power cord with the system turned On.
2.
Make sure that the power failure alarm comes on.
3.
Connect the power cable again.
4.
Make sure the alarm cancels.
Lock
before
use
2
Wa
rn
in
g
Kee
p u
pri
gh
t a
t a
ll t
im
es
Se
e m
an
ua
l fo
r
Op
era
tio
na
l I
nst
ru
cti
on
s.
Fac
tor
s a
ffe
ctin
g p
erf
orm
an
ce.
Se
rv
ice
an
d m
ain
te
na
nce
rec
om
me
nd
atio
ns
Ne
xt
ser
vic
e d
ate
!
U
se
O
n
ly
Is
of
lu
ra
n
e
3
4
5
1
.8
Lock
befo
re
use
Use Only
Enflurane
1
2
3
4
5
0%
.8 .6 .4 .2
War
nin
g
Keep upright at all times
Se
e m
an
ua
l fo
r
Ope
ration
al In
str
uc
tio
ns
.
Fac
tors
af
fect
ing
pe
rfor
ma
nc
e.
Se
rvic
e an
d m
ain
ten
an
ce
recom
men
da
tio
ns
Next ser
vice
dat
e
!
Of
Concentration
Control
AA.13.032
AA.43.005
Summary of Contents for 2 Vap. Trolley
Page 1: ...Aestiva 5 Operation Manual Part 1 Software Revision 4 X System Controls Operation Checkout ...
Page 6: ...Aestiva iv 1006 0938 000 iv ...
Page 14: ...Aestiva 1 8 1006 0938 000 1 8 ...
Page 36: ...Aestiva 2 22 1006 0938 000 ...
Page 76: ...Aestiva 4 4 1006 0938 000 ...
Page 102: ...Aestiva I 4 1006 0938 000 ...