3 Operation and Tutorial
1006-0938-000
3-35
3
Connecting Active AGSS
with a flow indicator
To use the optional active AGSS on a system which has a flow indicator (on
either the left or right side of the breathing system), connect it as follows.
Step 1
Connect the proper hose to the AGSS outlet connector
(f) on the rear of the base of the breathing system.
Step 2
Attach the other end to the hospital disposal system.
Note: To scavenge gas from a gas monitor, connect
tubing from the monitor to the 3.18 mm (1/8 inch)
hose barb (g) in the breathing system base.
Step 3
With the AGSS operating, verify that the flow indicator
ball (d) on the flow indicator (a) rises to the green zone
(c), indicating adequate flow.
Note: The ball in the upper red zone (b) indicates
excessively high extraction flow or a blocked filter. The
ball in the lower red zone (e) indicates extraction flow
rate is too low.
Step 4
Before you use the system, complete the Preoperative
Test procedure. Refer to the Appendix "Preoperative
Tests".
a
f
g
b
c
d
e
AB.23.174, AB.23.165
AB.23.162
AB.23.155