
Aestiva
3-20
1006-0938-000
Trigger Window
This control sets the range as a percent of the exhalation phase within which the
patient may trigger the next mechanical breath. Only active in SIMV and PSVPro
modes.
VT
This control sets the tidal volume delivered to the patient in the Volume Control
and SIMV modes.
Volume Control mode
The figure and table below show Volume Control settings.
Pressure Control mode
The figure and table below show Pressure Control settings.
SIMV mode
The figure and table below show SIMV settings.
Settings
V
T
Rate
I:E
Plimit
PEEP
Range
20 - 1500
4 - 100
2:1 - 1:8
12 - 100
Off, 4 - 30
Increments
varies
*
1 bpm
0.5
1 cmH2O
1 cmH2O
*
Increments of 1mL from 20 to 50, 5 mL from 50 to 100, 10 mL from 100 to 300, 25 mL from 300 to
1000, 50 mL from 1000 to 1500
Settings
Pinspired
Rate
I:E
Plimit
PEEP
Range
5 - 60
4 -1 00
2:1 - 1:8
12 - 100
Off, 4 - 30
Increments
1 cmH2O
1 bpm
0.5
1 cmH2O
1 cmH2O
AB.90.071
AB.90.025
AB.90.064
Summary of Contents for 2 Vap. Trolley
Page 1: ...Aestiva 5 Operation Manual Part 1 Software Revision 4 X System Controls Operation Checkout ...
Page 6: ...Aestiva iv 1006 0938 000 iv ...
Page 14: ...Aestiva 1 8 1006 0938 000 1 8 ...
Page 36: ...Aestiva 2 22 1006 0938 000 ...
Page 76: ...Aestiva 4 4 1006 0938 000 ...
Page 102: ...Aestiva I 4 1006 0938 000 ...