Aestiva
A-6
1006-0938-000
•Close the cylinder valve.
•Record the cylinder pressure after one minute. If the pressure decreases
more than 690 kPa (100 psig) there is a leak:
•Install a new cylinder gasket and tighten the tee handle as shown in
the Setup section of the Setup, maintenance and troubleshooting
manual.
•Do this step again. If the leak continues, do not use the system.
•Repeat step 3 for all cylinders.
•Close cylinder valves.
w
w
w
w
W
W
W
WA
A
A
AR
R
R
RN
N
N
NIIIIN
N
N
NG
G
G
G
Do not leave gas cylinder valves open if the pipeline supply is in
use. Cylinder supplies could be depleted, leaving an insufficient
reserve supply in case of pipeline failure.
4.
Connect the pipeline supplies.
5.
Turn on the system.
6.
Use the chart below to check pipeline pressure:
7.
Connect equipment to the pneumatic outlets as appropriate.
ANSI (USA and Intl.), Australian, Canadian, French,
Japanese
345 kPa (50 psig)
ISO, Italian, Scandinavian, South African, Spanish,
Swiss
414 kPa (60 psig)
Austrian, German
500 kPa (75 psig)