4
Preoperative Checklist
1006-0938-000
4-3
4-
Every time a different
clinician uses the system
Do a low-pressure leak test.
Before every patient
Look for damage, necessary drugs and equipment, correct breathing
circuit setup, and hazardous conditions.
Check vaporizer installation:
•Make sure the top of each vaporizer is horizontal (not on crooked).
•Make sure each vaporizer is locked and cannot be removed.
•Make sure the alarms and indicators operate correctly (Tec 6 vaporizer).
•Make sure you cannot turn on more than one vaporizer at the same time.
Do the breathing system tests:
•Make sure the one way valves and auxiliary equipment (humidifier, etc.)
operate correctly.
•With a circle breathing-circuit module, push the drain button for
≥
10 sec
to remove condensate.
•Ventilator circuit leak test.
•Bag/Manual circuit leak test.
• Bag/Manual circuit APL valve test.
•Circuit leak test.
Set the appropriate controls and alarm limits for the case.