3 Operation and Tutorial
1006-0938-000
3-21
3
PSVPro mode
PSVPro is pressure supported ventilation with apnea backup.
The figure below shows PSVPro settings. In this mode the Pinspired, Rate, and
Tinspired parameters are not active, but they may be changed.
During backup ventilation, the ventilator will ventilate the patient using the
SIMV-PC + PSV mode. All parameters shown are active in this mode.
Settings
V
T
Rate
Tinspired
Psupport
PEEP
Range
20 -1500
2 - 60
0.2 - 5.0
Off, 2-40
Off, 4 - 30
Increments
varies
*
1 bpm
0.1 s
1 cmH2O
1 cmH2O
*
Increments of 1mL from 20 to 50, 5 mL from 50 to 100, 10 mL from 100 to 300, 25 mL from 300 to
1000, 50 mL from 1000 to 1500
Settings
Pinspired
Rate
Tinspired
Psupport
PEEP
Range
5 - 60
2 - 60
0.2 - 5.0
Off, 2-40
Off, 4 - 30
Increments
1 cmH2O
1 bpm
0.1 s
1 cmH2O
1 cmH2O
AB.90.087
AB.90.086