2 System Controls and Menus
1006-0938-000
2-13
2
End Case key
End Case helps to prevent false alarms when no
patient is connected. It:
• Puts the apnea and volume alarms into Standby.
• Returns user selections to the most common
settings: Cardiac bypass off; Alarm limits shown; and
Heliox mode off.
• Sets the PEEP to 0 cmH
2
O (default value).
• Sets Plimit to one of two values: facility default or
4
4
4
40
0
0
0
cmH
2
O.
• Forces the circuit Leak Audio to On.
Mechanical ventilation must be off (set the Bag/Vent
switch to Bag or select the auxiliary common gas
outlet).
Item
Description
AB.29.006