Aestiva
3-22
1006-0938-000
Set inspiratory pause
(volume modes)
You can only use inspiratory pause in Volume Control or SIMV modes. In
pressure modes, Inspiratory Pause displays a message “No Pause
w
”.
When Pause is on, the inspiratory volume stays in the patient’s lungs for the set
pause time at the end of inspiration.
Pause can be set from Off to 60 percent of inspiratory time in increments of five
percent. Minimum pause time is 400 ms.
AB.90.092
Exp.
Insp.
Pause
Step 1
Select the Setup/Calibration menu.
• Push the menu key.
• Turn, then push the knob to select
the Setup/Calibration menu.
AB.29.0
1
3
AB.90.036
AB.90.03
7