LANcat
User Manual
Page 9-6
TWO-WAY SYSTEM CALIBRATION
To calibrate the remote unit on a TwoWay System:
1.
Connect the main unit to the CAT 3 calibration coupler,
using the modular plug Performance Module. (If you are
using the modular jack Performance Module, use the four-
foot CAT 5 patch cord).
2.
Connect remote unit to the open end of the CAT 3
calibration coupler using the Modular Plug Performance
module.
3.
Set the remote unit’s dial to Remote.
4.
Set the main unit’s dial to CAL.
5.
Highlight Remote in the Calibrate menu.
6.
When the test is finished, the screen returns to the Calibrate
menu.
7.
Turn the dial to Quick
. Press the UP or DOWN arrow
keys to highlight the CAT5 Basic Link test standard. Press
ENTER or SAVE (F4) to select the standard. The unit will
then perform a QUICK CHECK test.
8.
Once the test is complete, press VIEW (F3). Press the up or
down arrow key until the Attenuation results are displayed.
The Attenuation results should display values greater than
0.0 dB. If 0.0 dB is displayed for any or all pairs, perform
the Remote Calibration again.
NOTE
In order to ensure correct calibration, be sure
to switch the roles of both units, and repeat
the procedure above.