Chapter 9
Calibration
Page 9-3
USING MANUFACTURER’S CABLE SPECIFICATIONS
If you do not have a known length of cable available, but can
identify the cable manufacturer and part number, you may be able
to determine the rated cable NVP and load that NVP value into the
LANcat. Many common cable NVPs are listed in Appendix B,
Common NVP Values.
To load a known NVP value:
1.
Set the dial to CAL.
2.
Select NVP from the calibration menu and press CALIBRATE
(F4) or ENTER.
3.
Press INC (F3) or DEC (F2) to adjust the NVP value.
4.
Press ENTER to save the new NVP value. A message indicates
that the new value has been entered.
5.
Press ESC to exit the NVP calibration mode.
F1
F2
F3
F4
Entering a Known NVP Value