Vickers by Danfoss Target-Pro 2 Analyzer and Sampler Operation Manual V-PP-MC-0002-E

July 2023

110ml Bottle Sampler is designed
for use with the Target-Pro 2 Analyzer.
The bottle sampler
allows users to test fluids in situa-
tions where the analyzer cannot
be connected directly to the
hydraulic circuit. This helps over-
come limits of time and distance,
extending testing capabilities to
all equipment and locations. This
unit accepts both 100ml and
110ml bottles.

The bottle sampler offers the
option of de-aerating fluid sam-
ples before testing with the ana-
lyzer by applying a vacuum to the
chamber holding the sample bot-
tle. The de-aeration feature is only
needed when it is suspected, or
obvious, that air is present in the
fluid sample.

Two cap types are supplied with
the bottle sampler: a blank vacu-
um cap for use when de-aerating
fluid, and a pressure cap with an
anti-vacuum valve for use when
delivering a fluid sample to the
analyzer.

U s i n g

t h e

B o t t l e

S a m p l i n g

U n i t

1.

Clean the pump and hose provided with an appropriate
solvent, such as mineral spirits, before taking a sample.
Do not use acetone. Clean sample bottles and associated
products following the instructions on page 13.

2.

Draw at least

80ml of fluid

from the system into the bot-

tle provided. If necessary, use a hand pump.

3.

Insert the

waste fluid hose

in the waste disposable

bottle provided.

4.

Connect the waste fluid hose to the

waste connector

on

the Target-Pro 2. If you are using a hand pump, dis- charge
fluid from the hand pump into the

sample bottle

.

5.

Place the sample bottle into the

bottle sampling unit

.

6.

Connect the bottle sampling unit to the 12-volt DC

power supply

.

If the collected fluid contains air, follow steps 7–10 to de-
aerate the sample. Otherwise, skip to step 11.

Using the De-aeration Option

7.

Fit the

vacuum cap

to the bottle sampling unit.

8.

Switch off the bottle sampling unit, then turn the

selector knob

from Sample to Vacuum; the intermedi-

ate position has no function.

9.

Switch on the bottle sampling unit and leave it running
for several minutes, until all air bubbles have been
removed from the fluid sample.

10.

Switch off the bottle sampling unit; this automatically
vents the chamber to the atmosphere.

110ml Bottle Sampler
for Target-Pro 2

Summary of Contents for Vickers Target-Pro 2

Page 1: ...Vickers by Danfoss Target Pro 2 Analyzer and Sampler Operation Manual...

Page 2: ...and Sampler Operation Manual V PP MC 0002 E July 2023 Target Pro 2 Laser Particle Analyzer User Guide September 2004 Contents Operating Precautions 2 Introduction 3 ISO Cleanliness Code 3 Target Pro 2...

Page 3: ...d using mineral oil as the operating fluid The analyzer uses the light extinction principle in which two lasers are projected through the fluid and toward photo diodes Particles that interrupt the bea...

Page 4: ...5 to program the analyzer The display will show the Operations Screen To change analyzer settings follow this sequence 1 Test Reference Press 1 enter your reference details up to 16 characters and pre...

Page 5: ...previoustest and elimi nates cross contamination between samples 8 Press the flush valve push button to close the flush valve pushbutton light goes off Alternate method Proceed to step 9 pressing the...

Page 6: ...essel vented to the atmosphere To conserve battery life connect the analyzer to the power adapter for continuous operation Continuous Sampling Basic Operation 1 Follow instructions 1 5 on page 4 to se...

Page 7: ...ogging Press 4 once to log every test repeat press 4 again to deselect logging Note If you deselect log every test the analyzer will only store the test results when the Target Cleanliness Level is ac...

Page 8: ...n appropriate solvent such as mineral spirits before taking a sample Do not use acetone Clean sample bottles and associated products following the instructions on page 13 2 Draw at least 80ml of fluid...

Page 9: ...a minimum 1 Half fill the sampling bottle with the fluid to be analyzed 2 Re cap the bottle and shake it vigorously for 30 seconds 3 Discard the fluid following current COSHH legislation 4 Repeat step...

Page 10: ...from the PC menu 6 When the transfer is complete switch the analyzer off Software Installation The internal battery can sustain 8 hours of continuous oper ation approximately 100 tests after charging...

Page 11: ...press the flush valve and see if fluid passes into the waste disposal bottle Body text Note If you suspect high water aeration levels contact Danfoss for advice Warning If excessive system contaminat...

Page 12: ...roscope counting examines the particles differently from APCs the code is given with two scale numbers only These are at 5 m and 15 m equivalent to the 6 m c and 14 m c of the APCs ISO 4406 Cleanlines...

Page 13: ...ers These relate to the use of the correct viscosity mineral fluid An even cleaner level may be needed if the operation is severe such as high frequency fluctuations in loading high temperature or hig...

Page 14: ...000 25 23 17 100 000 CorrelationbetweenParticleSizesObtainedusingACFTDandNIST CalibrationMethods Particle Size Obtained Using ACFTD SIZE ISO 4402 1991 ISOMTD NIST ISO 11171 size m m c 1 4 2 2 4 6 3 5...

Page 15: ...required by ISO 9000 QS9000 and similar quality management sys tems A procedure for determining the performance of APCs means the user can set minimum acceptablelevels Benefits include improved calib...

Page 16: ...Phone 45 7488 2222 Danfoss Power Solutions US Company 2800 East 13th Street Ames IA 50010 USA Phone 1 515 239 6000 Danfoss Power Solutions Trading Shanghai Co Ltd Building 22 No 1000 Jin Hai Rd Jin Q...

Reviews: