A n a l y z e r U p p e r
C o n t a m i n a t i o n L i m i t
The analyzer upper operating
limit is set at 24/22/20.
Tests that result in particle
counts exceeding any scale
number in the three-part ISO
upper limit has the scale
number replaced by an
asterisk. Associated particle
counts on the printout are
replaced by X’s.
See example.
Further Test—Same
Sampling Point
To repeat a test on the same
sample point, press the
Start button [1]. The test
number will increase auto-
matically.
Further Test—Different
Sampling Point/Same
System
Repeat steps
7–12 above.
To change test reference/test
mode data, repeat steps
1–12 on pages 4–5.
Further Test—New System
Repeat steps 1 through 12
on pages 4–5.
Online—Normal
Particle count and ISO Code to 4406
standard(NASCode1638displayed)
Online—Dynamic
ISO and NAS codes, complete with aver-
age analysis
Online—Normal
Particle counts displayed; NAS Code 1638
standard(ISOCode4406displayed)
Running a Test
Ta k i n g a S a m p l e —
N o r m a l , D y n a m i c ,
Tr i p l e / B o t t l e , S h o r t
7.
Press the flush valve push-
button to open the flush
valve; pushbutton lights up
to indicate valve is open.
• Leave valve open for at
least one minute or 200ml
of fluid; this removes
trapped air and fluid from
the previous test and elimi-
nates cross-contamination
between samples.
8.
Press the flush valve push-
button to close the flush
valve; pushbutton light
goes off.
• Alternate method: Proceed
to step 9—pressing the
start button automatically
closes the flush valve
before sampling begins.
9.
Press the start
button [1]. The
analyzer will
begin the sam- pling cycle.
10.
Results will be automati-
cally displayed on the
screen and, if Auto Print
mode is on, will be auto-
matically printed at the
end of the sampling cycle.
(If Auto Print is off, press
the Print key for a copy of
the printed results.)
11.
After displaying the
results, the analyzer auto-
matically discharges the
sample fluid to waste.
Test status is shown as
Emptying.
12.
When the Sampling and
Emptying cycles are com-
pleted, the test status is
shown as Idle.
Results are automatically
stored to memory. To
download, follow the
instructions on page 9.
Shutting Down
1. Press red button to switch
off analyzer.
2. Disconnect minimess con-
nector on sampling hose
to isolate fluid supply.
3. Remove fluid sampling
hose from analyzer.
4. Remove waste fluid hose
from analyzer.
5. Replace end capso n
sampling hose; wipe
clean and store.
6. Connect waste fluid hose
quick-coupling end fittings
together; wipe clean and
store.
I n t e r p r e t i n g R e s u l t s
See page 11 for hydraulic
component manufacturers’
recommendations on stan-
dard cleanliness require-
ments for various applica-
tions.
ISO 4406 and NAS 1638 are not directly
comparative. Please refer to page 11.