Vickers by Danfoss Target-Pro 2 Analyzer and Sampler Operation Manual V-PP-MC-0002-E

July 2023

To change thermal printer paper, remove the four thumb-
screws securing the cover and serrated paper cutter.
Thermal printer paper is sensitized on one side only, and
must be fed into the printer mechanism as shown below.

Using a finger, press down on the print head spring at point
A; tilt the print head to the open position by pulling it back at
point B.

Feed the paper under the roller and pull the paper end out
of the mechanism. Return the print head back to its normal
position by pressing on the green lever at point C. Make
sure the print head spring returns to the position shown
above.

Important:

Do not operate the printer without paper; doing

so will damage the printer. Replace the roll when the “end
of roll” indication appears on the paper.

Warranty and Recalibration

The Target-Pro 2 is guaranteed for one year after delivery.
Danfoss recommends you return the analyzer once each
year for recalibration.

Printer Paper

Troubleshooting

Specifications

P r o b l e m C h e c k

1.

LCD screen remains blank after switching on

• Make sure analyzer battery is charged, or has

been charging for at least 15 minutes

• See if LED illuminates when power adapter is

connected to DC power input socket on analyzer

2.

Unexpected results obtained from sample

• Check the fluid sampling hose connections

at both the system and analyzer ends

• Confirm that there is a free flow of fluid to the

analyzer: press the flush valve and see if fluid
passes into the waste disposal bottle Body
text:

Note:

If you suspect high water/aeration levels, contact

Danfoss for advice.

Warning:

If excessive system contamination is suspected,

flush the analyzer using the bottle sampling unit in conjunc-
tion with a suitable solvent, such as mineral spirits.

Do not use acetone.

Technology

Automatic Optical Particle Analyzer

Laser Package

Twin Laser and Twin Optical Diode
Detectors

LCD display

(backlit)

Sensitivity

>4, 6, 14, 21, 25, 38, 50, 68 µm(c), Micron
range to revised ISO 4406 Standard

Accuracy/Repeatability

Better than 3% typical

Calibration

Each unit is individually calibrated with
ISO Medium Test Dust (MTD) based on
ISO 11171:1999 on equipment certified by
I.F.T.S.

Analysis Range

ISO 8 to ISO 24 to ISO 4406
(NAS 1638-2 to 12)

Report/PrintFormat

ISO and NAS codes, with individual
particle counts as a built-in option
Printer Fixed head thermal printer
(384 dots per line)

Target-Pro

15 ml. (normal), 30 ml. (dynamic)

2 SampleVolume

24 ml. (bottle sampler)
15 ml. (continuous), 8ml. (short)

Operation

Max. system working pressure, 400 bar;
Min. working pressure, 2 bar

Viscosity Range

to 400 centistokes

OperatingTemperature

+5 to +80°C

Fluid Compatibility

Mineral oil and petroleum based fluids
(consult Danfoss for other fluids)

Typical Test Time

Result in <2.5 mins. (normal test)

Power

Internal rechargeable battery
(AC charger) or external 12/24 volt DC
power supply

Data Storage

600 tests

Computer Interface

RS 232 (serial) communication port

Hose Connections

minimess fittings
microbore hose, 1.5 m waste fluid hose

Dimensions

Height 210mm; Depth 260mm;
Width 430mm; Weight 7.6 kg

Danfoss reserves the right to alter these specifications without notice.

Summary of Contents for Vickers Target-Pro 2

Page 1: ...Vickers by Danfoss Target Pro 2 Analyzer and Sampler Operation Manual...

Page 2: ...and Sampler Operation Manual V PP MC 0002 E July 2023 Target Pro 2 Laser Particle Analyzer User Guide September 2004 Contents Operating Precautions 2 Introduction 3 ISO Cleanliness Code 3 Target Pro 2...

Page 3: ...d using mineral oil as the operating fluid The analyzer uses the light extinction principle in which two lasers are projected through the fluid and toward photo diodes Particles that interrupt the bea...

Page 4: ...5 to program the analyzer The display will show the Operations Screen To change analyzer settings follow this sequence 1 Test Reference Press 1 enter your reference details up to 16 characters and pre...

Page 5: ...previoustest and elimi nates cross contamination between samples 8 Press the flush valve push button to close the flush valve pushbutton light goes off Alternate method Proceed to step 9 pressing the...

Page 6: ...essel vented to the atmosphere To conserve battery life connect the analyzer to the power adapter for continuous operation Continuous Sampling Basic Operation 1 Follow instructions 1 5 on page 4 to se...

Page 7: ...ogging Press 4 once to log every test repeat press 4 again to deselect logging Note If you deselect log every test the analyzer will only store the test results when the Target Cleanliness Level is ac...

Page 8: ...n appropriate solvent such as mineral spirits before taking a sample Do not use acetone Clean sample bottles and associated products following the instructions on page 13 2 Draw at least 80ml of fluid...

Page 9: ...a minimum 1 Half fill the sampling bottle with the fluid to be analyzed 2 Re cap the bottle and shake it vigorously for 30 seconds 3 Discard the fluid following current COSHH legislation 4 Repeat step...

Page 10: ...from the PC menu 6 When the transfer is complete switch the analyzer off Software Installation The internal battery can sustain 8 hours of continuous oper ation approximately 100 tests after charging...

Page 11: ...press the flush valve and see if fluid passes into the waste disposal bottle Body text Note If you suspect high water aeration levels contact Danfoss for advice Warning If excessive system contaminat...

Page 12: ...roscope counting examines the particles differently from APCs the code is given with two scale numbers only These are at 5 m and 15 m equivalent to the 6 m c and 14 m c of the APCs ISO 4406 Cleanlines...

Page 13: ...ers These relate to the use of the correct viscosity mineral fluid An even cleaner level may be needed if the operation is severe such as high frequency fluctuations in loading high temperature or hig...

Page 14: ...000 25 23 17 100 000 CorrelationbetweenParticleSizesObtainedusingACFTDandNIST CalibrationMethods Particle Size Obtained Using ACFTD SIZE ISO 4402 1991 ISOMTD NIST ISO 11171 size m m c 1 4 2 2 4 6 3 5...

Page 15: ...required by ISO 9000 QS9000 and similar quality management sys tems A procedure for determining the performance of APCs means the user can set minimum acceptablelevels Benefits include improved calib...

Page 16: ...Phone 45 7488 2222 Danfoss Power Solutions US Company 2800 East 13th Street Ames IA 50010 USA Phone 1 515 239 6000 Danfoss Power Solutions Trading Shanghai Co Ltd Building 22 No 1000 Jin Hai Rd Jin Q...

Reviews: