O p e r a t i n g P r e c a u t i o n s
Battery
• The Target-Pro 2 uses a nickel-cadmium (NiCd) battery.
Charge it for at least 24 hours before first use.
• For optimum performance, fully discharge and recharge
the battery once a month: Leave the Target-Pro 2
switched on until the LCD screen goes blank, then switch
off the Target-Pro 2, connect the charging adapter, and
allow it to charge the battery for 24 hours.
• To prolong battery life, switch off the analyzer when it is
not in use.
Internal Cleaning
Do not
clean the Target-Pro 2 or Bottle Sampler with ace-
tone or other solvents that are not compatible with nitrile
seals. The recommended cleaning fluid for internal flushing
is mineral spirits.
2
Vickers by Danfoss
Target-Pro 2 Analyzer and Sampler Operation Manual V-PP-MC-0002-E July 2023
Target-Pro 2
Laser Particle Analyzer
User Guide:
September 2004
C o n t e n t s
Operating Precautions
2
Introduction
3
ISO Cleanliness Code
3
Target-Pro 2 Operation
3
Key and Connector Diagrams
3
Online Operation
3
Analyzer Set-up
4
Test Set-up
4
Running a Test
5
Shutting Down
5
Continuous Sampling
6–7
External W iring Details
8
Memor y Recall
8
Battery Charging / Printer Paper
9
Software Installation
9
ISO 4406 Cleanliness Code System
10
NAS 1638 Cleanliness Code System
11
Hydraulic Component Manufacturers Recommendations
11
Hydraulic System Target Cleanliness Levels
1 1
110ml Bottle Sampler for Target-Pro 2
12–13
Appendix
13–14
Specifications and Part Numbers
1 5
Troubleshooting
15
Danfoss reserves the right to alter specifications without notice. Except as permitted by such license,
no part of this publication may be reproduced without prior written permission of Danfoss.