background image

1 4

Vickers by Danfoss Target-Pro 2 Analyzer and Sampler Operation Manual V-PP-MC-0002-E

July 2023

Other Standards

Although ISO 4406 is used extensively in the hydraulics
industry, other standards are occasionally required. Direct
comparison is not always possible, but this table
provides some estimates.

Approximate Equivalents of Contamination Classes

ISO 4406

DEF.STD 05/42 [7]

NAS 1638[5]

SAE 749[8]

Table A

Table B

ISO 11218[6]

13/11/08

2

14/12/09

3

0

15/13/10

4

1

16/14/09

400F

16/14/11

5

2

17/15/09

400

17/15/10

800F

17/15/12

6

3

18/16/10

800

18/16/11

1,300F

18/16/13

7

4

19/17/11

1,300

2,000F

19/17/14

8

5

20/18/12

2,000

20/18/13

4,400F

20/18/15

9

6

21/19/13

4,400

6,300F

21/19/16

10

22/20/13

6,300

22/20/17

11

23/12/14

15,000

23/21/18

12

24/22/15

21,000

25/23/17

100,000

Correlation between Particle Sizes Obtained using ACFTD and NIST
Calibration Methods

Particle Size Obtained Using
ACFTD SIZE (ISO 4402:1991)

ISOMTD NIST (ISO 11171) size

µm

µm(c)

1

4.2

2

4.6

3

5.1

4

5.8

5

6.4

6

7.1

7

7.7

8

8.4

9

9.1

10

9.8

11

10.6

12

11.3

13

12.1

14

12.9

15

13.6

16

14.4

17

15.2

18

15.9

19

16.7

20

17.5

21

18.2

22

19.0

23

19.7

24

20.5

25

21.2

26

22.0

27

22.7

28

23.5

29

24.2

30

24.9

31

25.7

32

26.4

33

27.1

34

27.9

35

28.5

36

29.2

37

29.9

38

30.5

39

31.1

40

31.7

TheTarget-Pro 2 is calibrated with ISO MediumTest Dust (to ISO 11171).
The correlation between particle sizes and the ACFTD (old standard) to
the ISOMTD (new standard) is as follows :

ACFTD

<1

5

15

25

30

50*

75*

100*

ISOMTD

4

6

14

21

25

38

50

68

* To be confirmed by NIST.

This table is only a guideline. The exact relationship between ACFTD sizes and the NIST
sizes may vary from instrument to instrument, depending on the characteristics of the parti-
cle counter and original ACFTD calibration.

CleanlinessCodes
(continued)

Summary of Contents for Vickers Target-Pro 2

Page 1: ...Vickers by Danfoss Target Pro 2 Analyzer and Sampler Operation Manual...

Page 2: ...and Sampler Operation Manual V PP MC 0002 E July 2023 Target Pro 2 Laser Particle Analyzer User Guide September 2004 Contents Operating Precautions 2 Introduction 3 ISO Cleanliness Code 3 Target Pro 2...

Page 3: ...d using mineral oil as the operating fluid The analyzer uses the light extinction principle in which two lasers are projected through the fluid and toward photo diodes Particles that interrupt the bea...

Page 4: ...5 to program the analyzer The display will show the Operations Screen To change analyzer settings follow this sequence 1 Test Reference Press 1 enter your reference details up to 16 characters and pre...

Page 5: ...previoustest and elimi nates cross contamination between samples 8 Press the flush valve push button to close the flush valve pushbutton light goes off Alternate method Proceed to step 9 pressing the...

Page 6: ...essel vented to the atmosphere To conserve battery life connect the analyzer to the power adapter for continuous operation Continuous Sampling Basic Operation 1 Follow instructions 1 5 on page 4 to se...

Page 7: ...ogging Press 4 once to log every test repeat press 4 again to deselect logging Note If you deselect log every test the analyzer will only store the test results when the Target Cleanliness Level is ac...

Page 8: ...n appropriate solvent such as mineral spirits before taking a sample Do not use acetone Clean sample bottles and associated products following the instructions on page 13 2 Draw at least 80ml of fluid...

Page 9: ...a minimum 1 Half fill the sampling bottle with the fluid to be analyzed 2 Re cap the bottle and shake it vigorously for 30 seconds 3 Discard the fluid following current COSHH legislation 4 Repeat step...

Page 10: ...from the PC menu 6 When the transfer is complete switch the analyzer off Software Installation The internal battery can sustain 8 hours of continuous oper ation approximately 100 tests after charging...

Page 11: ...press the flush valve and see if fluid passes into the waste disposal bottle Body text Note If you suspect high water aeration levels contact Danfoss for advice Warning If excessive system contaminat...

Page 12: ...roscope counting examines the particles differently from APCs the code is given with two scale numbers only These are at 5 m and 15 m equivalent to the 6 m c and 14 m c of the APCs ISO 4406 Cleanlines...

Page 13: ...ers These relate to the use of the correct viscosity mineral fluid An even cleaner level may be needed if the operation is severe such as high frequency fluctuations in loading high temperature or hig...

Page 14: ...000 25 23 17 100 000 CorrelationbetweenParticleSizesObtainedusingACFTDandNIST CalibrationMethods Particle Size Obtained Using ACFTD SIZE ISO 4402 1991 ISOMTD NIST ISO 11171 size m m c 1 4 2 2 4 6 3 5...

Page 15: ...required by ISO 9000 QS9000 and similar quality management sys tems A procedure for determining the performance of APCs means the user can set minimum acceptablelevels Benefits include improved calib...

Page 16: ...Phone 45 7488 2222 Danfoss Power Solutions US Company 2800 East 13th Street Ames IA 50010 USA Phone 1 515 239 6000 Danfoss Power Solutions Trading Shanghai Co Ltd Building 22 No 1000 Jin Hai Rd Jin Q...

Reviews: