Appendix

New ISO Standard Test Dust and its
effect on ISO Contamination Control Standards

For more than 25 years, the measurement of solid particles
in hydraulic fluids hasdepended on AC Fine Test Dust
(ACFTD), a product manufactured by General Motors.
ACFTD was used for calibrating Automatic Particle Counters
(APCs), testing components such as oil filters, and testing
the contaminant sensitivity of hydraulic components.

The particle size distribution of ACFTD, determined using an
optical microscope, formed the basisof ISO 4402—the
method for calibrating APCs. The limitations of optical micro-
scopes meant that distribution was uncertain at particle
sizes smaller than 5 microns. Consistency between batches
was also a regular concern.

After General Motors decided to stop making ACFTD, the
International Organization for Standardization (ISO) defined
requirements for a replacement, and asked the National
Institute of Standards and Technology (NIST ) to produce a
standard, traceable reference substance. The new dust’s
particle size distribution has been accurately determined
with the aid of scanning electron microscopes and image
analysis techniques.

New Test Dust Benefits

The new ISO Medium Test Dust (ISOMTD) uses similar
materials to ACFTD, but in a slightly coarser grade to reduce
particle counting errors. ISOMTD is produced to a standard
distribution and stringent quality control procedures for reli-
able consistency. In addition, production of ISOMTD
includes a revised ISO APC calibration method.

The result is a controlled reference test dust with uniform
particle size distribution. It provides the traceability required
by ISO 9000, QS9000 and similar quality management sys-
tems. A procedure for determining the performance of
APCs means the user can set minimum acceptablelevels.
Benefits include improved calibration techniques and proce-
dures, greater particle count reproducibility with different
equipment, and more accurate and consistent filter test
results.

Effect on Industry

T H E

I N T R O D U C T I O N

O F

I S O M T D

H A S

R E Q U I R E D

C H A N G E S

C E R T A I N

I S O

S T A N D A R D S .

T H E S E

I N C L U D E :

ISO 4402

1991

Hydraulic fluid power Calibration
of liquid automatic

particle counters

ISO 4406

1987

Hydraulic fluid power
Code for defining the level of

contamination by solid particles

ISO 4572

1981

Hydraulic fluid power Filters
Multi-pass method for evaluating filtration

performance of filter elements

To prevent confusion over which standards to use, ISO is
updating 4402 to ISO 11171, and 4572 to ISO 16889. ISO
standards written for the new test dust will use a new iden-
tifier, ‘(c)’. Micron sizes according to ISO 11171 will be
expressed as ‘µm(c)’ and Beta ratios according to ISO
16889 will be expressed as ‘Bx(c)’,e.g. ‘B5(c)’.

In the new ISO 4406, new calibration sizes are used to give
the same cleanliness codes as the ‘old’ calibration sizes of 5
and 15 microns. This will prevent the need to change sys-
tem cleanliness specifications.

The proposed cleanliness codes for APCs will be formed from
three particle counts at 4, 6 and 14 microns, with 6 and 14
microns corresponding closely to the previous 5 and
15 micron measurements. Users may still quote just two
counts of 6 microns and 14 microns for APCs.

Counts derived by microscope counting methods are not
affected, so the particle sizes used for microscopy, at 5 and
15 microns, will remain unchanged. The only real effect
users will experience is improved particle count accuracy;
there will be no change in filter performance or ISO cleanli-
ness levels.

Summary of Contents for Vickers Target-Pro 2

Page 1: ...Vickers by Danfoss Target Pro 2 Analyzer and Sampler Operation Manual...

Page 2: ...and Sampler Operation Manual V PP MC 0002 E July 2023 Target Pro 2 Laser Particle Analyzer User Guide September 2004 Contents Operating Precautions 2 Introduction 3 ISO Cleanliness Code 3 Target Pro 2...

Page 3: ...d using mineral oil as the operating fluid The analyzer uses the light extinction principle in which two lasers are projected through the fluid and toward photo diodes Particles that interrupt the bea...

Page 4: ...5 to program the analyzer The display will show the Operations Screen To change analyzer settings follow this sequence 1 Test Reference Press 1 enter your reference details up to 16 characters and pre...

Page 5: ...previoustest and elimi nates cross contamination between samples 8 Press the flush valve push button to close the flush valve pushbutton light goes off Alternate method Proceed to step 9 pressing the...

Page 6: ...essel vented to the atmosphere To conserve battery life connect the analyzer to the power adapter for continuous operation Continuous Sampling Basic Operation 1 Follow instructions 1 5 on page 4 to se...

Page 7: ...ogging Press 4 once to log every test repeat press 4 again to deselect logging Note If you deselect log every test the analyzer will only store the test results when the Target Cleanliness Level is ac...

Page 8: ...n appropriate solvent such as mineral spirits before taking a sample Do not use acetone Clean sample bottles and associated products following the instructions on page 13 2 Draw at least 80ml of fluid...

Page 9: ...a minimum 1 Half fill the sampling bottle with the fluid to be analyzed 2 Re cap the bottle and shake it vigorously for 30 seconds 3 Discard the fluid following current COSHH legislation 4 Repeat step...

Page 10: ...from the PC menu 6 When the transfer is complete switch the analyzer off Software Installation The internal battery can sustain 8 hours of continuous oper ation approximately 100 tests after charging...

Page 11: ...press the flush valve and see if fluid passes into the waste disposal bottle Body text Note If you suspect high water aeration levels contact Danfoss for advice Warning If excessive system contaminat...

Page 12: ...roscope counting examines the particles differently from APCs the code is given with two scale numbers only These are at 5 m and 15 m equivalent to the 6 m c and 14 m c of the APCs ISO 4406 Cleanlines...

Page 13: ...ers These relate to the use of the correct viscosity mineral fluid An even cleaner level may be needed if the operation is severe such as high frequency fluctuations in loading high temperature or hig...

Page 14: ...000 25 23 17 100 000 CorrelationbetweenParticleSizesObtainedusingACFTDandNIST CalibrationMethods Particle Size Obtained Using ACFTD SIZE ISO 4402 1991 ISOMTD NIST ISO 11171 size m m c 1 4 2 2 4 6 3 5...

Page 15: ...required by ISO 9000 QS9000 and similar quality management sys tems A procedure for determining the performance of APCs means the user can set minimum acceptablelevels Benefits include improved calib...

Page 16: ...Phone 45 7488 2222 Danfoss Power Solutions US Company 2800 East 13th Street Ames IA 50010 USA Phone 1 515 239 6000 Danfoss Power Solutions Trading Shanghai Co Ltd Building 22 No 1000 Jin Hai Rd Jin Q...

Reviews: