Radiographic image scanner EAGLE.PS
21
EQUIPMENT OPERATION
Erasing the image in the Phosphorous plate
After the image is Retrieved, the Phosphorous plates are automatically erased. However, it is suggested
that the phosphorous plate is erased manually when:
The image Retrieved exceeds the 24 hour period.
The image was not properly Retrieved and there was an error during the erasure of the phosphorous plate.
In case of doubt whether the phosphorous plate is clean or not.
To erase a phosphorous plate through the Software, choose the option
Erase Plate PSPL and proceed with the insertion of phosphorous plate into the
equipment.
To erase a phosphorous plate through the equipment, press the "Erase" button
located in the center of the main panel. Wait for the White LED to remain constant
to insert the phosphorous plate.
EAGLE.PS DICOM viewer operation - System Execution
The EAGLE.PS DICOM Viewer executable is located in the directory where the image was exported in the
portable folder with executable option for Windows and Mac.
View Image DICOM
The option to open a DICOM file is available in the Toolbar
Figure 1 - Toolbar EAGLE.PS DICOM Viewer
The user can select from 1 (one) to 10 images to be opened simultaneously in the EAGLE.PS DICOM
Viewer
Figure 2 - EAGLE.PS Main Screen DICOM Viewer