DEVICE SPECIFIC INFORMATION 

UV Series  

 

Available Lamps Types & Number of Lamps:  

 

Numerals = Number of Lamps 

N/A = Not Available  

 

UVA 

Narrow Band 

UVB 

Broad Band 

UVB 

UVA / Narrow  

Band UVB 

12 12 

12 

N/A 

16 16 

16 

N/A 

24 

24 

24 

12 + 12 

 

This section covers topics that are specific to the UV Series.  

The Flex Dosimetry or Flex Timer operating instructions are covered on pages sixteen (16) through 
twenty-two (22).   

 

UV Series Unpacking & Assembly 

The UV Series full-body device is shipped upright in a large cardboard container.  Exercise caution as 
the fluorescent lamps can be broken if the unit is not unpacked properly.   

Using as many people as available but at least two, move the box into the area where the device will 
be used.  If there are no stairs, and since the unit is on castors, it may be easier to remove it from the 
box at the earliest opportunity so that you can simply roll it into position.  If there are stairs, you may 
consider hiring a professional moving company to move the device into position. 

To unpack the unit, first use metal cutters to cut the banding that encircles the package.  Use extreme 
caution as these bands are under tension and can snap outward when the tension is released.  Next, 
remove the front cardboard cover.  Use a heavy-duty screwdriver to remove the staples.   Again be 
careful as the staples are sharp and can cause injury.   When the front cardboard cover has been 
removed, carefully pull the device forward on its castors until it is clear of the box.  Take care not to 
discard the accessory package which contains your protective eyewear. 

Once in position, remove the shipping bars.  A ¾” end wrench will be required to loosen and remove 
the bolts that hold the shipping bars in place.  There is no further assembly required other than to plug 
the unit into the appropriate electrical outlet (See Electrical Outlet Requirements below).  If your 
device is equipped with an optional platform, place the platform inside the machine once it is located 
in its final position.   

Circuit Breaker 

Each UV Series unit is equipped with a circuit breaker that is 
located under the  frame, at the back right hand side of the 
machine.  Its location is indicated by the circuit breaker label 
(right).  When setting up your unit, ensure that the breaker is in 
the ON position.    

UV Series Electrical Requirements 

The number of lamps in a unit determines the voltage, amperage, wire size and outlet necessary to 
safely support the unit’s operation.  The table and diagrams below define these requirements for each 

   

MAIN DISCONNECT 

    CIRCUIT BREAKER 

 

 l 

=

        

=

 

ON = RIGHT 

 

=

        

OFF

 = 

LEFT

 

  

UV Series

 

Summary of Contents for 7 Series

Page 1: ...OPERATING INSTRUCTIONS AND GENERAL INFORMATION HOME PHOTOTHERAPY MANU0027 Rev 12 1 17 11 7 Series 2 Series UV Series M Series 1 Series...

Page 2: ...able Lamp Types Number of Lamps 13 M Series Unpacking Assembly Guide 13 Special Instructions for Combination Units 13 M Series Electrical Outlet Requirements 14 How to Position Yourself 14 M Series La...

Page 3: ...aintenance Schedule 23 Cleaning 23 Lamp Replacement 23 Calibration 24 Refilling a Light Prescription 24 Warranty 25 How Daavlin Handles Warranty and Service 26 Responsibilities of the Buyer 26 Contact...

Page 4: ...t detailing all symbols located on your device and their meaning DANGEROUS VOLTAGE ATTENTION CONSULT ACCOMPANYING DOCUMENTS NON IONIZING RADIATION TYPE B EQUIPMENT I ON OFF EARTH ground PROTECTIVE EAR...

Page 5: ...xample shown below Please read the label carefully as it contains important safety information for you In addition to the warning label an identification label indicates the serial number and date of...

Page 6: ...ormation on the UV 7 2 M and 1 Series devices The next section beginning on page sixteen 16 Flex Operating Instructions provides operating instructions for all devices Finally the last section Care of...

Page 7: ...any area where water or moisture might collect and should be protected from access by children Exposure to ultraviolet light over extended periods of time may cause carpets wall coverings and furnish...

Page 8: ...not resume the treatment until the redness has subsided If redness and discomfort are severe you should discuss this with your physician Increases of less than 15 seconds may be necessary for some pat...

Page 9: ...eleased Next remove the front cardboard cover Use a heavy duty screwdriver to remove the staples Again be careful as the staples are sharp and can cause injury When the front cardboard cover has been...

Page 10: ...to any one panel than to another Standing closer to one of the sides may cause localized streaking or burning UV Series Lamp Removal and Replacement 1 Consult the Lamp Replacement Guide for replacemen...

Page 11: ...mbly of your device choose a place close to where the unit will be installed to unpack it however if there are stairs or narrow hallways it may be easier to move the device after it has been assembled...

Page 12: ...the device still lying on its side stand the device upright With one person lifting at the top and a second person at the foot to catch the unit gently lift and tip the device forward until both feet...

Page 13: ...vlin service to determine an appropriate replacement How to Position Yourself The recommended distance from the lamps for treatment is 9 inches 22cm Mark that distance on the floor in some manner A si...

Page 14: ...0 hertz units are shipped with a grounded three conductor wall plug Units designed for other voltages may be shipped with no wall plug which would then have to be locally obtained and installed Hood L...

Page 15: ...six 6 for replacement lamp specifications 2 Unplug the machine and disconnect the power cord for the hood before removing the hood 3 Remove the hood from the base The hood lamps are then easily acces...

Page 16: ...ld be as close as possible to the center of the fixture for best results Larger Area Treatment Stand the device so that the lamps are perpendicular to the surface on which the device is standing Posit...

Page 17: ...is available with two types of control system Flex Dosimetry and Flex Timer Verify the type of control system you have and proceed to the applicable operating instructions Note If your serial number...

Page 18: ...ed at the factory are shown on the Power Output Certificate included in your manual Note It is not necessary to take a power output reading before each treatment We suggest once every couple of weeks...

Page 19: ...g your hands elbows or forearm for example directly on the plastic lamp guard the energy output there is higher When you take a power output reading this is the value that is shown on the display All...

Page 20: ...dose on the display alternating with the elapsed time If the complete dose is not delivered meaning that the backup time was reached before the dose was fully delivered this is an indication that the...

Page 21: ...ian s protocol Another method is to use the factory power output levels shown on the Power Output Certificate to determine estimated time equivalents needed to deliver doses in milliJoules UVB or Joul...

Page 22: ...are different than the UVB output levels Each chart is clearly marked UVA Blue Light Red Light or UVB Entering a treatment time 1 Press until the first digit of the desired time is displayed If the t...

Page 23: ...using the same treatment time Press and release both keys at the same time Enter Special Notes The maximum time that can be entered is 59 minutes and 59 seconds It is not possible to enter 60 seconds...

Page 24: ...lamp areas 7 2 and UV Series the lamp area can be cleaned by carefully vacuuming with the grid installed Occasionally it may be desirable to remove the lamps and grid or plastic lamp guard on M and 1...

Page 25: ...ure limiting software are shipped from the factory with a forty exposure prescription Each time the lamps are turned on the number of available exposures is reduced by one When the machine is in sleep...

Page 26: ...or altered outside the factory in any way so as to affect the design or which has been subject to misuse negligence or accident or operated in any way other than in accordance with our operating inst...

Page 27: ...avlin technicians or subcontractors which falls outside the warranty will be billed at the prevailing service rate at the time of service Responsibilities of the Buyer In the event of a malfunction th...

Reviews: