GENERAL OVERVIEW 

 
This manual covers the following phototherapy equipment: 
 
  

UV Series 

  

7 Series 

  

2 Series 

  

M Series 

  

1 Series 

   

 

 
Your device may be equipped with a variety of different lamps and may be manufactured to a number 
of different electrical specifications to meet national and international standards. It may also be 
equipped with either a Flex™ integrating dosimetry control or a Flex timer control.  And, in either case, 
your physician may have prescribed your device with the FlexRx system that limits the total number of 
exposures that the device will deliver.  This manual is applicable for all lamp configurations, electrical 
specifications and control types.  
 
This section of the manual, the 

General Overview, 

covers topics such as site location, the initial 

inspection of the device, lamp type confirmation, eye protection, precautions and phototherapy in 
general.    These topics are common to all devices.    
 
Beginning on page eight (8) you will find 

Device Specific Information

 on the UV, 7, 2, M, and 1 

Series devices.   
 
The next section, beginning on page sixteen (16), 

Flex Operating Instructions 

provides operating 

instructions for all devices.  
 
Finally, the last section, 

Care of Your Unit, 

beginning on page twenty-three (23), contains information 

concerning the maintenance, cleaning, re-lamping, and calibration (where applicable) of your device 
along with warranty and service contact details. 
 
We suggest that you first read the sections concerning site selection and general inspection under 

Overview 

and then the instructions that are 

Specific

 to your device with regard to unpacking and 

assembly. 
 
When you first receive the device: 

Inspect the box  

Any signs of shipping damage should be noted on the shipping receipt that you will be asked to sign 
by the delivering carrier.  If damage is discovered after unpacking the unit, save all packing materials 
and call Daavlin at 1-800-322-8546 for inspection and repair.   

            The delivering carrier must be notified of any shipping damage within twenty-four (24) hours 

   

 to protect your right to an insurance claim. 

Locate the following materials 

 

 

   Time chart     

    Power Output Certificate   

   Protective Eyewear   

 

   Manual  

 
 

Summary of Contents for 7 Series

Page 1: ...OPERATING INSTRUCTIONS AND GENERAL INFORMATION HOME PHOTOTHERAPY MANU0027 Rev 12 1 17 11 7 Series 2 Series UV Series M Series 1 Series...

Page 2: ...able Lamp Types Number of Lamps 13 M Series Unpacking Assembly Guide 13 Special Instructions for Combination Units 13 M Series Electrical Outlet Requirements 14 How to Position Yourself 14 M Series La...

Page 3: ...aintenance Schedule 23 Cleaning 23 Lamp Replacement 23 Calibration 24 Refilling a Light Prescription 24 Warranty 25 How Daavlin Handles Warranty and Service 26 Responsibilities of the Buyer 26 Contact...

Page 4: ...t detailing all symbols located on your device and their meaning DANGEROUS VOLTAGE ATTENTION CONSULT ACCOMPANYING DOCUMENTS NON IONIZING RADIATION TYPE B EQUIPMENT I ON OFF EARTH ground PROTECTIVE EAR...

Page 5: ...xample shown below Please read the label carefully as it contains important safety information for you In addition to the warning label an identification label indicates the serial number and date of...

Page 6: ...ormation on the UV 7 2 M and 1 Series devices The next section beginning on page sixteen 16 Flex Operating Instructions provides operating instructions for all devices Finally the last section Care of...

Page 7: ...any area where water or moisture might collect and should be protected from access by children Exposure to ultraviolet light over extended periods of time may cause carpets wall coverings and furnish...

Page 8: ...not resume the treatment until the redness has subsided If redness and discomfort are severe you should discuss this with your physician Increases of less than 15 seconds may be necessary for some pat...

Page 9: ...eleased Next remove the front cardboard cover Use a heavy duty screwdriver to remove the staples Again be careful as the staples are sharp and can cause injury When the front cardboard cover has been...

Page 10: ...to any one panel than to another Standing closer to one of the sides may cause localized streaking or burning UV Series Lamp Removal and Replacement 1 Consult the Lamp Replacement Guide for replacemen...

Page 11: ...mbly of your device choose a place close to where the unit will be installed to unpack it however if there are stairs or narrow hallways it may be easier to move the device after it has been assembled...

Page 12: ...the device still lying on its side stand the device upright With one person lifting at the top and a second person at the foot to catch the unit gently lift and tip the device forward until both feet...

Page 13: ...vlin service to determine an appropriate replacement How to Position Yourself The recommended distance from the lamps for treatment is 9 inches 22cm Mark that distance on the floor in some manner A si...

Page 14: ...0 hertz units are shipped with a grounded three conductor wall plug Units designed for other voltages may be shipped with no wall plug which would then have to be locally obtained and installed Hood L...

Page 15: ...six 6 for replacement lamp specifications 2 Unplug the machine and disconnect the power cord for the hood before removing the hood 3 Remove the hood from the base The hood lamps are then easily acces...

Page 16: ...ld be as close as possible to the center of the fixture for best results Larger Area Treatment Stand the device so that the lamps are perpendicular to the surface on which the device is standing Posit...

Page 17: ...is available with two types of control system Flex Dosimetry and Flex Timer Verify the type of control system you have and proceed to the applicable operating instructions Note If your serial number...

Page 18: ...ed at the factory are shown on the Power Output Certificate included in your manual Note It is not necessary to take a power output reading before each treatment We suggest once every couple of weeks...

Page 19: ...g your hands elbows or forearm for example directly on the plastic lamp guard the energy output there is higher When you take a power output reading this is the value that is shown on the display All...

Page 20: ...dose on the display alternating with the elapsed time If the complete dose is not delivered meaning that the backup time was reached before the dose was fully delivered this is an indication that the...

Page 21: ...ian s protocol Another method is to use the factory power output levels shown on the Power Output Certificate to determine estimated time equivalents needed to deliver doses in milliJoules UVB or Joul...

Page 22: ...are different than the UVB output levels Each chart is clearly marked UVA Blue Light Red Light or UVB Entering a treatment time 1 Press until the first digit of the desired time is displayed If the t...

Page 23: ...using the same treatment time Press and release both keys at the same time Enter Special Notes The maximum time that can be entered is 59 minutes and 59 seconds It is not possible to enter 60 seconds...

Page 24: ...lamp areas 7 2 and UV Series the lamp area can be cleaned by carefully vacuuming with the grid installed Occasionally it may be desirable to remove the lamps and grid or plastic lamp guard on M and 1...

Page 25: ...ure limiting software are shipped from the factory with a forty exposure prescription Each time the lamps are turned on the number of available exposures is reduced by one When the machine is in sleep...

Page 26: ...or altered outside the factory in any way so as to affect the design or which has been subject to misuse negligence or accident or operated in any way other than in accordance with our operating inst...

Page 27: ...avlin technicians or subcontractors which falls outside the warranty will be billed at the prevailing service rate at the time of service Responsibilities of the Buyer In the event of a malfunction th...