![background image](http://html1.mh-extra.com/html/daavlin/7-series/7-series_operating-instructions-manual_3057306013.webp)
12
11. Insert the perpendicular grid wires into the two holes in the top lamp plate and push the grid all the
way up. Swing the bottom of the grid in toward the lamps and insert the grid wires into the two holes
in the bottom lamp plate. Then, lower the grid until the bracket touches the bottom lamp plate. Using
a ¼” hex driver/socket, screw the bracket to the bottom lamp plate using one of the hex head, self-
tapping screws that are provided.
7 and 2 Series Electrical Requirements
Both the 7 and 2 Series devices are shipped within the USA and Canada with a standard three prong
grounded wall plug. Units designed for other voltages may be shipped with no wall plug, which would
then have to be locally obtained and installed. All 7 and 2 Series devices can be plugged into a
standard, grounded wall outlet.
Both 7 and 2 Series devices are equipped with onboard fuses. The fuse holders are located just
below where the power cord plugs into the unit. Should an onboard fuse blow, call Daavlin service to
determine an appropriate replacement.
How to Position Yourself
The recommended distance from the lamps for treatment is 9 inches (22cm). Mark that distance on
the floor in some manner. A similar system should be devised if you are using a 2 Series on a
tabletop or other non-floor surface area.
Your torso should not be closer to the lamps than this line. Standing closer to the lamps will cause
localized streaking or burning. Greater distances will increase the time necessary to receive a
therapeutic dose.
7 Series & 2 Series Lamp Removal & Replacement
1. Consult the Lamp Replacement Guide for replacement lamp specifications.
2. Unplug the machine. Then, using a ¼” hex driver/socket, remove the screw from the bottom
of the grid that is in front of the lamp(s) to be removed.
3. Lift the grid out of the holes in the bottom lamp plate, pull the bottom of the grid forward, then
slide it out of the holes in the top of the unit and set it aside.
4. Grasp the lamp to be removed with both hands and press down until it clears the top socket,
then remove the lamp. Reverse the process to install lamps and grid.
5. Reset the lamp hours age to zero. See instructions on page twenty-two (22).