19
6. Press
▲
until the number “2” is displayed.
7. Press
◄
to move the “2” one space to the left.
8. Press
▲
until the number “8” is displayed.
9. Press
◄
to move the “8” one space to the left and repeat until all desired digits are shown on
the display. In this example the last digit is a zero, so once you have moved the “8” into
position, the “0” already shows and you can move on to step 10.
10. When the desired dose is displayed, press ENTER.
Put on your goggles and position yourself per the instructions of your specific device.
11. Begin the treatment by again pressing ENTER.
To repeat the same dose for another exposure, press the END key to clear the display, and then
press the MODE and POWER keys at the same time and release. This displays and enters the
previous dose. Press ENTER to start the timer again.
Note:
The system beeps three times to signal the end of the treatment and then flashes the
delivered dose on the display alternating with the elapsed time. If the complete dose is not
delivered, meaning that the backup time was reached before the dose was fully delivered, this is
an indication that the backup times being entered are not long enough. Verify that the sensor is
clean, unobstructed and centered in the hole in the reflector. Then, re-take your power output
reading and adjust the back-up time. It is normal to have to increase back-up times due to aging
lamps and/or because of slow starting lamps, low room temperatures or low line voltage.
Special 1 Series Instructions when the device is positioned away from your body:
When setting up a treatment where it is impractical for the area being treated to be in direct contact
with the plastic guard, you should position the device 8 inches (20cm) from the body. Since the power
output there is 1/3 lower than at the plastic lamp guard, it is necessary to multiply the prescribed dose
by three. Therefore, if your dose was 280 milliJoules, you would have to enter a value of 840
milliJoules along with a backup time for 840 milliJoules.