16
OPERATING THE FLEX CONTROL SYSTEM
Unlocking the Device
To prevent unauthorized use, the device will “self lock” when it has not been used for twenty (20)
minutes. To unlock, press any key. The word “CODE” will appear. Press
▲
until the numeral 7
appears, then press ENTER. This is the factory default key code. If you wish to personalize or
change your keycode, see the next section.
Setting or Changing the Key Code:
Press MODE,
◄
and ENTER at the same time
Then…
Press
▲
until the display reads 9
Then…
Press ENTER
Then…
Press MODE and
▲
at the same time
Then…
Enter desired Key Code (Set to 0000 to disable)
Then…
Press ENTER
Flex Dosimetry or Flex Time Control
Daavlin phototherapy equipment is available with two types of control system; Flex Dosimetry and
Flex Timer. Verify the type of control system you have and proceed to the applicable operating
instructions.
Note
:
If your serial number contains ( ****-** X -****) then the unit has a dosimetry controller.
If your serial number contains ( ****-** T -****) then the unit has a timer only controller.
Flex Dosimetry Instructions
Daavlin integrating dosimetry systems have a built-in power
meter that measures energy output. This information is used to
determine the exact amount of time that each treatment will last.
Unlike timer based systems, the dose is entered in milliJoules
(UVB) or Joules (UVA) (Blue Light).
The output of fluorescent lamps varies with age, use and other
factors such as room temperature. Measuring for these
variations and then integrating them into the dose time is an
effective way to ensure the precision of each treatment.