Function Menu
Spectra II
44
Operating Manual
08.17
6.5 Interface
6.5.1 COM1
Off:
serial interface Off
On (mode 1):
serial interface On
On (mode 2):
serial interface On; no error message appears in case
of a transmission error
Indication of bits which are transferred per second (speed of data
transfer).
Value range: 1200, 2400, 4800, 9600, 19200, 38400, 57600 and
115200.
None:
No parity
Even:
Even parity
Odd:
Odd parity
Setting of data bits.
Value range: 7 or 8 Bits.
Indication of stop bits between bytes.
Value range: 1 or 2 stop bits.
6.5.2 General Parameters
SOH:
Start of data transfer block
Hex format 01
ETB:
End of data transfer block
Hex formal 17
Two different start / en signs can be set. The settings are normally
SOH = 01 HEX and ETB = 17 HEX. Several host computers cannot
process these signs and therefore SOH = 5E HEX and ETB = 5F
cannot be set.
Off:
After starting a print order no more data is received.
Standard:
After starting a print order the printer buffer receives data
as long as it is filled.
Advanced:
During a current print order data is received and
processed.
Check whether the data are transferred via the interface.
COM1 mode
Baud rate
Parity
Data bits
Stop bits
SOH/ETB
Data memory
Port test