PN 4237523A
B-1
B
B
DIFF COMPARISON
B.1
INTRODUCTION
You can perform manual differentials as a measure of good QC practice or as recommended
by your laboratory, state, and federal protocols.
At the end of this appendix are blank log sheets used in the procedure.
B.2
PROCEDURE
1.
Collect two different normal blood samples. Follow NCCLS H3-A2 procedures.
The samples must be:
r
Collected in K
3
EDTA with 1.5 mg/mL of anticoagulant to blood
r
Maintained at room temperature
r
Within the reference range of:
NE 40 to 72%
LY 17 to 45%
MO 4 to 12%
EO 0 to 10%
BA 0 to1%
r
Analyzed between 30 minutes and 5 hours after draw.
2.
Label the samples.
Draw
1
2
3
4
5
6
7
8
9
10
11
12
1