PN 4237523A
3-6
OPERATION PRINCIPLES
SAMPLE FLOW
2.
The baths drain into the waste chamber (Figure 3.6).
Figure 3.6 Baths Draining
3.
Figure 3.7 applies to steps 3 through 7. The BSV rotates to segment three portions of the
sample:
r
1.6 µL for the RBC bath dilution
r
28 µL for the WBC bath dilution
r
28 µL for the WBC diff.
4.
The air pump transfers the diff portion to the sample line.
5.
Dispensers send diluent to the BSV to pick up and dilute the RBC and WBC portions.
6.
The CBC lytic reagent pumps send the CBC lytic reagent (LYSE S III or LYSE S 4) to the
WBC bath. It lyses the RBCs and converts the hemoglobin into a stable compound.
7.
Mixing bubbles enter both baths.
Hgb
lamp
Hgb
preamp
card
WBC
bath
Waste
chamber
RBC
bath
Summary of Contents for HmX Hematology Analyzer
Page 4: ...PN 4237523A iv REVISION STATUS ...
Page 14: ...PN 4237523A xiv CONTENTS ...
Page 18: ...PN 4237523A xviii INTRODUCTION COMPUTER PROGRAM STATEMENT ...
Page 28: ...PN 4237523A 1 10 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Page 34: ...PN 4237523A 2 6 INSTALLATION INTERUNIT CONNECTIONS ...
Page 52: ...PN 4237523A 3 18 OPERATION PRINCIPLES PARAMETERS AND THEIR DERIVATION ...
Page 70: ...PN 4237523A 4 18 SPECIFICATIONS CHARACTERISTICS INTERFERING SUBSTANCES ...
Page 78: ...PN 4237523A 6 4 REPORTING OPTIONS GRAPHIC FORMAT ...
Page 84: ...PN 4237523A 7 6 BAR CODE SPECIFICATIONS BAR CODE DECODER ...
Page 98: ......
Page 110: ...PN 4237523A 8 GLOSSARY GLOSSARY ...
Page 124: ...PN 4237523A 14 INDEX INDEX ...
Page 126: ......