PN 4237523A
2-3
INSTALLATION
INTERUNIT CONNECTIONS
2
Date Format
Choose which of the following date formats fits your location:
r
European(DD/MM/YY)
r
Japanese(YY/MM/DD)
r
U.S. (MM/DD/YY).
Your Beckman Coulter Representative installs your selection at installation.
2.3
INTERUNIT CONNECTIONS
Reagent and Waste Connections
Figure 2.1 shows the tubing connections from the instrument to the reagent and to the waste
containers. The pak lyse, pak preservative, and lytic reagent are placed at the correct level on
the counter beside the unit.
Be sure to keep all reagent lines isolated from anything electrical.
CAUTION
Possible reagent siphoning effect and priming problems could occur if a reagent container is
placed above the level of the Diluter. Place the reagent containers at the same level as the system or below.
IMPORTANT
Possible incorrect platelet counts can be reported due to bubbles in the line, and reagent
priming problems could occur if the reagent tubing lengths are too long. Do not exceed 8 ft from the rear of
the system to any reagent container.
Summary of Contents for HmX Hematology Analyzer
Page 4: ...PN 4237523A iv REVISION STATUS ...
Page 14: ...PN 4237523A xiv CONTENTS ...
Page 18: ...PN 4237523A xviii INTRODUCTION COMPUTER PROGRAM STATEMENT ...
Page 28: ...PN 4237523A 1 10 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Page 34: ...PN 4237523A 2 6 INSTALLATION INTERUNIT CONNECTIONS ...
Page 52: ...PN 4237523A 3 18 OPERATION PRINCIPLES PARAMETERS AND THEIR DERIVATION ...
Page 70: ...PN 4237523A 4 18 SPECIFICATIONS CHARACTERISTICS INTERFERING SUBSTANCES ...
Page 78: ...PN 4237523A 6 4 REPORTING OPTIONS GRAPHIC FORMAT ...
Page 84: ...PN 4237523A 7 6 BAR CODE SPECIFICATIONS BAR CODE DECODER ...
Page 98: ......
Page 110: ...PN 4237523A 8 GLOSSARY GLOSSARY ...
Page 124: ...PN 4237523A 14 INDEX INDEX ...
Page 126: ......