OPERATION PRINCIPLES
COUNTING AND SIZING
3
Sweep Flow
The sweep flow is a steady stream of diluent that flows behind the RBC aperture during the
sensing periods (B). This keeps cells from swirling back into the sensing zone. Because these
swirling cells (A) would be peripherally sensed, their pulse height would be similar to Plt
pulse heights. See Figure 3.11.
Figure 3.11 Sweep Flow
Pulse Editing
When cells pass through the aperture near the edge or at an angle rather than at the center,
they create atypical pulses. Pulse editing technology eliminates atypical pulses from the
analysis because they distort the true size of the cell. This prevents atypical pulses from
influencing size measurements.
RBC Count and Size Distribution
During RBC sensing, pulses that represent cells from 36 fL to 360 fL are classified as RBCs
and are sorted by size into 256 channels to build the RBC histogram. Using a system of
moving averages, the computer smooths the histogram curve.
Plt Count and Size Distribution
During RBC sensing, pulses that represent cells from 2 to 20 fL are classified as platelets. The
system automatically extends the sensing time if platelet accumulation is below a
predetermined level. The system sorts the Plt pulses by size into 64 channels to build the Plt
histogram.
Sensing
zone
Swirling
effect
No
sweep
flow
A
B
Cell
Sweep
flow
Cell
Sensing
zone
To
waste
Summary of Contents for HmX Hematology Analyzer
Page 4: ...PN 4237523A iv REVISION STATUS ...
Page 14: ...PN 4237523A xiv CONTENTS ...
Page 18: ...PN 4237523A xviii INTRODUCTION COMPUTER PROGRAM STATEMENT ...
Page 28: ...PN 4237523A 1 10 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Page 34: ...PN 4237523A 2 6 INSTALLATION INTERUNIT CONNECTIONS ...
Page 52: ...PN 4237523A 3 18 OPERATION PRINCIPLES PARAMETERS AND THEIR DERIVATION ...
Page 70: ...PN 4237523A 4 18 SPECIFICATIONS CHARACTERISTICS INTERFERING SUBSTANCES ...
Page 78: ...PN 4237523A 6 4 REPORTING OPTIONS GRAPHIC FORMAT ...
Page 84: ...PN 4237523A 7 6 BAR CODE SPECIFICATIONS BAR CODE DECODER ...
Page 98: ......
Page 110: ...PN 4237523A 8 GLOSSARY GLOSSARY ...
Page 124: ...PN 4237523A 14 INDEX INDEX ...
Page 126: ......